27
Guidance and manufacturer’s
declaration-Electromagnetic immunity
Immunity test
Compliance level
Electromagnetic
environment -
guidance
Electrostatic
discharge
(ESD) IEC
61000 - 4-2
Compliant
+/- 8kV Contact,
+/ 2kV,
+/- 4kV, +/- 8kV,
+/- 15kV Air
Electrical
fast transient/
burst IEC
61000-4-4
Compliant
+/-2 kV for power
supply lines
+/-1kV for input/
output lines
Surge IEC
61000-4-5
Compliant
+/-1 kV
differential mode
+/-2kV common
mode
Voltage dips.
Short
Interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
Compliant
<5% Ut (>95%
dip in Ut) for 0.5
cycle, 40% Ut
(60% dipinUt)
for 5 cycles ,
70% Ut (30% dip
in Ut)
for 25 cycles,
<5% Ut
(>95% dip in Ut)
for 5 cycles
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Compliant
30 A/m
Conducted RF
IEC 61000-4-6
Compliant
3 Vrms 150kHz
to 80 MHz,
6 Vrms at ISM
band
Radiated RF
IEC 61000-4-3
Compliant
3 V/m 80MHz to
2.7GHz
ClearUP® is intended for use in the
electromagnetic environment specified below.
The customer or the user of ClearUP should
assure that it is used in such an environment.