metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site 
survey, 

a

 should be less than the compliance level in each frequency

range.

b

Interference may occur in the vicinity of equipment marked with the
following symbol:

NOTE1  At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE2    These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by  absorption  and  reflection  from  structures,

objects and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and 
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the FT-2020 is used exceeds
the applicable RF compliance level above, the FT-2020 should be observed to verify normal operation. If abnormal performance is observed, 
additional measures may be necessary, such as re-orienting or relocating the FT-2020.

Recommended separation distances between

portable and mobile RF communications equipment and the FT-2020

The FT-2020 is intended for use in an electromagnetic environment (for home healthcare) in which radiated RF disturbances are controlled. The 
customer or the user of the FT-2020 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile 
RF communications equipment (transmitters) and the FT-2020 as recommended below, according to the maximum output power of the 
communications equipment.

Rated maximum output power

of transmitter

W

Separation distance according to frequency of transmitter

M

150 kHz to 80 MHz

d = 

1,2 

80 MHz to 800 MHz

d = 

1,2 

800 MHz to 2,7 GHz

d = 

2,3 

0,01

N/A

0,12

0,23

0,1

N/A

0,38

0,73

1

N/A

1,2

2,3

10

N/A

3,8

7,3

100

N/A

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance

d

in meters (m) can be estimated using the equation

applicable  to  the  frequency  of  the  transmitter,  where

P

is  the  maximum  output  power  rating  of  the  transmitter  in  watts  (W)  according  to  the  transmitter

manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and 
people.

Manufacturer’s declaration-electromagnetic immunity

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment

The FT-2020 is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the FT-2020 
should assure that it is used in such an environment.

Test

frequency

(MHz)

Band

a)

(MHz)

Service

a)

Modulation

b)

Maximum  power

(W)

Distance

(m)

IMMUNITY  TEST

LEVEL

(V/m)

Compliance

LEVEL

(V/m)

(for home

healthcare)

385

380–390

TETRA  400

Pulse

modulation  b)

18  Hz

1,8

0,3

27

27

450

430–470

GMRS  460,

FRS  460

FM  c)

 ±5  kHz  deviation

1 kHz  sine

2

0,3

28

28

www.tiptemp.com