Symbol Descriptions
Caution
Please read the instructions for use
Battery Recycling
BF type applied part
Paper Recycling
Stand-by
IP22
Classification for water ingress and
particulate matter.
Manufacturer’s declaration-electromagnetic emissions
The FT-2020
is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the FT-2020
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
(for home healthcare environment)
RF emissions
CISPR 11
Group 1
The FT-2020 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The FT-2020 is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Manufacturer’s declaration – electromagnetic immunity
The FT-2020
is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the FT-2020
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
–
guidance
(for home healthcare environment)
Electrostatic
discharge(ESD) IEC 61000-
4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8 kV,
±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at
least 30%
Power frequency(50, 60 Hz)
magnetic field IEC 61000-4-
8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
The FT-2020 power frequency magnetic fields should be at levels
characteristic of a typical location in a typical home healthcare
environment.
Manufacturer’s declaration – electromagnetic immunity
The FT-2020 is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the FT-2020
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment
–
guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
Recommended separation distance:
Error: Reference source not found
d = 1,2 Error: Reference source not found
80MHz to 800 MHz
d = 2,3 Error: Reference source not found 800MHz to 2,7 GHz
Where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer and
d
is the recommended separation distance in
www.tiptemp.com
TIP TEMPerature Products – 340 W Broad Street, Burlington, NJ 08016 – 1-800-TIP-TEMP – [email protected]