background image

Ref No.

032020

Symbol Descriptions

Caution

Please read the instructions for use

Battery Recycling

BF type applied part

Paper Recycling

Stand-by

IP22

Classification for water ingress and 
particulate matter.

Manufacturer’s declaration-electromagnetic emissions

The FT-2020

 

is intended for use in the electromagnetic environment (for home healthcare) specified below.The customer or the user of the FT-2020 

should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment 

– 

guidance

(for home healthcare environment)

RF emissions

CISPR 11

Group 1

The FT-2020 uses RF energy only for its internal function. Therefore, its RF emissions are 
very low and are not likely to cause any interference in nearby electronic equipment. 

RF emissions

CISPR 11

Class B

The FT-2020 is suitable for use in all establishments, including domestic establishments and 
those directly connected to the public low-voltage power supply network that supplies 
buildings used for domestic purposes.

Manufacturer’s declaration – electromagnetic immunity

The FT-2020

 

is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the FT-2020 

should assure that it is used in such an environment.
Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment 

– 

guidance

(for home healthcare environment)

Electrostatic 
discharge(ESD) IEC 61000-
4-2

Contact:±8 kV 
Air±2 kV,±4 kV,±8 kV,
±15 kV

Contact:±8 kV 
Air±2 kV,±4 kV,±8 
kV,±15 Kv

Floors should be wood, concrete or ceramic tile. If floors are 
covered with synthetic material, the relative humidity should be at 
least 30%

Power frequency(50, 60 Hz) 
magnetic field IEC 61000-4-
8

30 A/m
50 Hz or 60 Hz

30 A/m
50 Hz and 60 Hz

The FT-2020 power frequency magnetic fields should be at levels 
characteristic of a typical location in a typical home healthcare 
environment.

Manufacturer’s declaration – electromagnetic immunity

The FT-2020 is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of the FT-2020 
should assure that it is used in such an environment.
Immunity test

IEC 60601 test level

Compliance

level

Electromagnetic environment 

– 

guidance

(for home healthcare environment)

Radiated RF 
IEC 61000-4-3

10 V/m
80 MHz – 2,7 GHz 
80 % AM at 1 kHz

10 V/m
80 MHz – 2,7 
GHz 
80 % AM at 1 
kHz 

Recommended separation distance:

d = 1,2 

Error: Reference source not found

d = 1,2 Error: Reference source not found

  

80MHz to 800 MHz

d = 2,3 Error: Reference source not found  800MHz to 2,7 GHz 
Where

 P

 is the maximum output power rating of the transmitter in watts (W) according 

to the transmitter manufacturer and 

is the recommended separation distance in 

www.tiptemp.com