Display
Cause
Remedy
Err.7
System error
Send in for repair
Err.11
Value could not be
Temperature outside
calculated
the admissible
temperature
measuring range
Calibration
Display
Cause
Remedy
CAL.-Err.1
Value was inad-
Clean electrode,
missible, cell
re-calibrate
constant too high:
If error repeated
electrode defective
-> replace electrode
Buffer solution
Use fresh buffer
unusable
solution
CAL.-Err.2
Value was
Replace electrode
inadmissible, cell
constant too low:
electrode defective
Buffer solution
Use fresh buffer
unusable
solution
CAL.-Err.4
Wrong temperature
Calibration only
during calibration
possible in range
from 10 … 40
°
C