10.2. Checking retention parts 

 

 

 

All maintenance operations must be done by skilled personnel! 

Periodically (at least every 6 months), make sure all the retention parts are tight (screws, pins, nuts, 
etc.) and check the condition of joints. 

 

 

Periodically (at least every 6 months), it have to be done the service 
check by authorised servicer.  All this check – services have to be 
noted in service book with date and sign of service done. Only with 
the signed service book the warranty is valid.

   

 

10.3. Electrical diagram         

 

Electrical scheme:

   

 

 

 
 

 

 

A  central box CBJ 2 -Linak 

CBJ2 

B  battery BAJ 1 -Linak 

BAJ1 

C  manual control HB 81 -Linak 

J00724 

D  actuator LA 44 -Linak 

442100 

E  cord motor -Linak 

914573 

H  Battery Charger CHJ 2 -Linak 

CHJ2 

 
Connecting the battery on a wall charger (CHJ2) and on box (CBJ 2).                                                                     
Battery is inserted into CHJ2 or CBj 2  to place it on the bus and pushed it into place. 
Remove the battery from CHJ2 or CBJ 2 by holding the handle at the back of the battery and then press 
it to remove the battery. 
 

8.3.1. Charging the battery                                                                                                 

 

How do we know when we have to connect the battery charger? When the voltage drops below a 
certain point we hear a sound signal, which informs us that it is necessary to connect the battery 
charger. 
 
Wall battery charger has two signal lights (yellow and green). Green light shows when the charger is 
connected to power. Yellow light is turn on when the battery is charging when the battery is fully 
charged, the yellow light turns off. Charging the battery takes about 4 hours.  

 
  

 

Before first use the battery have to be connected 12 hours! 
 

 

Do not charging the battery in bathroom! 

Summary of Contents for 155040

Page 1: ...unted shower stretcher art 155040 and Electric wall mounted changing table art 155140 is a medical device in compliance with Directive 93 42 EGS and technical standard CE EN 60601 1 CE EN 60601 2 52 Timo d o o Ajdovščina Vipavska cesta 4i 5270 Ajdovščina Tel 00386 0 5 3689 330 Fax 00386 0 5 3663805 e mail info timo si http www timo si Carefully read these instructions before instalation ...

Page 2: ...4 6 3 Transport 14 7 INSTALATION 14 7 1 Wall mounting 15 7 2 Assembly 15 8 CONTROLS 16 8 1 Use 17 8 2 Manual Controls 17 8 2 1 Backrest section movement art 155040 17 8 2 2 Side rails movement 18 8 2 3 Positioning of mattress platform into vertical position 18 8 3 Electrical regulations 19 8 3 1 Raising and lowering the mattress platform 19 9 CLEANING AND DEZINFECTION 20 9 1 Cleaning 20 9 2 Desinf...

Page 3: ...f the main operations to be performed by operators with specific experience acquired for instance by attending training courses on the manufacturer s premises This manual should be considered an integral part of the equipment and must be looked after until this is scrapped In case of loss ask the manufacturer for a new copy Make sure all the users have fully understood how to use the equipment and...

Page 4: ...his manual GENERIC HAZARD IMPORTANT Abide by the ins tructions in the follo wing paragraphs OPERATOR TYPE A Equipment user OPERATOR TYPE B Equipment maintenance technician IN CONFORMITY WITH DIRECTIVE 93 42 EC DATE OF MANUFACTURE ...

Page 5: ... covered by a 1 year warranty starting on the day of shipment The warranty does not cover component parts covered by a separate original manufacturer s warranty The warranty is valid as and from the date of shipment of the equipment The warranty shall only be considered valid if the equipment has been used correctly following the instructions provided in this manual Faulty parts will be repaired o...

Page 6: ...ss platform into vertical position against the wall The Electric wall mounted changing table art 155140 has been planned and created for the treatment and changing of patient particularly in every day care use Entirely made of epoxy coated steal with a special cataphoresis anti corrosion treatment following criteria of safety for the patient and the operator The Electric wall mounted changing tabl...

Page 7: ...pecial mechanism for lowering and lifting flexible drain hose mattress and pillow Description of The Electric wall mounted changing table art 155140 TIMO d o o Ajdovščina has produced The Electric wall mounted changing table art 155140 for the changing which is designed to care and changing immobile persons patients The Electric wall mounted changing table art 155140 is in size and design made by ...

Page 8: ...hower stretcher art 155040 and Electric wall mounted changing table art 155140 the operator must be perfectly aware of how the Electric wall mounted shower stretcher art 155040 and Electric wall mounted changing table art 155140 works of where the controls are located and of all technical functional features Such awareness can be acquired by carefully reading this manual The nursing staff must pro...

Page 9: ...ic wall mounted shower stretcher art 155040 and Electric wall mounted changing table art 155140 Any alterations to the equipment not expressly authorised by TIMO d o o Ajdovščina shall relieve the latter of all civil and criminal liability Removing any safety or protection devices on the appliance is strictly forbidden periodically at weekly intervals check the integrity of such devices and make s...

Page 10: ... moving parts of the mattress platform and the fixed wall stretchers structure outstanding risk of being crushed During wall stretchers positioning be careful not to crush people or things between the head end and the walls against which the wall stretchers is rested outstanding risk of crushing ACCIDENTAL FALLING When getting in and out of the wall stretchers make sure the height of the stretcher...

Page 11: ...in Chap 8 9 This person should be perfectly aware of all the precautions and operating instructions indicated in this manual The operator is forbidden from performing operations other than those indicated in Chap 8 9 The MAINTENANCE TECHNICIAN is the person appointed to transport install start adjust clean repair change the equipment and perform maintenance on the wall stretcher Such person must b...

Page 12: ... B Serial no C Safe working load D Max lifting load E Description type F Symbol CE 3 Name of producer 5 TEHNICAL DETAILS The products of the company TIMO d o o Ajdovščina are subject to continous upgrading and consequently the technical specifications of the product may be changed without prior notice 5 1 Wall mounted shower stretcher and changing table specification Dimension and degree art 15504...

Page 13: ... Charger 100 240 V 50 60 Hz 650mA Weight of the wall stretcher 100 kg 6 HANDLING 6 1 Lifting from the package Lifting from the package must only be done by skilled personnel trained to perform this type of operation The Customer is responsible for obtaining the equipment suitable for lifting from the package the wall stretcher Before lifting from the package the wall stretcher make sure no one is ...

Page 14: ...anging table art 155140 manufactured by TIMO d o o Ajdovščina are shipped to customers in two parts the wall fitting panel and mattress platform lying part The wall fitting pannel have to be fitted to a wall as described in that instruction in chapter 7 Than you have to fit the mattress platform lying part as described in that instruction in chapter 7 Transport is generally done by truck The tempe...

Page 15: ... to be load bearing strong enough to carry a 150 kg load on each of the wall ficings screw fi 10x100 The wall area is required to be at least 220 cm wide and 160 cm hight flat plumb and clear of any obstacles The plywood support must be in contact with and supported by the floor The wall fitting panel cannot be fitted to metal frame or similar type stud wall unless it has been fully reinforced wit...

Page 16: ...oes not have his her arms or legs located in any opening for example in the protection side rail Before using some of the functions the Electric wall mounted shower stretcher art 155040 and Electric wall mounted changing table art 155140 height lowering backrest lowering make sure that nothing and no one especially children is underneath the mattress platform During raising and lowering movements ...

Page 17: ...tion can be quickly lowered by means of the handle underneath the mattress platform on the front side 8 2 1 1 Raising the backrest section 1 Grip the backrest part 2 Pull the handle at the side of the mattress platform upwards without forcing fig 12 fig 12 3 Rise up the platform as far as desired position 4 First release the handle and then the backrest 8 2 1 2 Lowering the backrest section art 15...

Page 18: ...always check that it is properly locked in place fig 13 Lowering the side rails Firmly grip the top part of the side rail and lift the part that is marked with the red arrow as shown fig 14 and with the other hand squeezed the trigger lever together and release the safety side until the lower position is reached fig 14 8 2 3 Positioning of mattress platform into vertical position The manual regula...

Page 19: ... lowering of mattress platform All movements can be performed for a maximum time of 2 minutes After this time a thermal cut out will interrupt movements until resetting Expected rest period approx 18 min Movements will start again automatically after this time ATTENTION Always tell the patient to get in and out of the wall stretcher with the mattress platform in fully lowered position 8 3 1 Liftin...

Page 20: ... not contain phosphates phosphorus or formaldehyde 9 2 Disinfection The wall stretcher disinfection must be done using suitable non corrosive products Disinfectants must not contain phosphates phosphorus or formaldehyde and must have a pH between 6 and 8 NOTE The temperature of the washing water must not be above 85 C 10 MAINTENANCE All maintenance operations must be done by skilled personnel 10 1...

Page 21: ...CHJ 2 Linak CHJ2 Connecting the battery on a wall charger CHJ2 and on box CBJ 2 Battery is inserted into CHJ2 or CBj 2 to place it on the bus and pushed it into place Remove the battery from CHJ2 or CBJ 2 by holding the handle at the back of the battery and then press it to remove the battery 8 3 1 Charging the battery How do we know when we have to connect the battery charger When the voltage dro...

Page 22: ... raised 1 It is possible that the battery is completely flat or defective or is not properly connected 2 It is possible that the hand control panel is defective 3 It is possible that the actuator is defective 4 It is possible that the feeding cable is defective 5 It is possible that the control box is defective 1 Connect battery to a charger or call a technician to replace the battery 2 Call a tec...

Page 23: ...Screw imbus M8 x 80 6 Adjustable feet ø50 M10x50 22 Washers pvc ø40 10x4 7 Wall cover 2 23 Screw imbus M10 x 20 8 Wall cover 24 Slider 20 9 Wall cover D 25 Plugin 25x25 10 Wall cover L 26 Support for cover 11 screw M10 x 80 27 Screw imbus M8 x 70 12 screw M10 x 60 28 Control box CBJ2 13 screw M10 x 55 29 screw ø 4x15 14 screw M10 x 60 30 Battery linak baj 15 Self tapping nut M10 31 Spring washer D...

Page 24: ...NUMBER 1 nosilni okvir enodelne stenske preg mize 2 ležišče 1700x700 3 safety side 4 plastic end cap ø 20 5 plastic end cap 50x30 6 plastic end cap 25x25 7 screw M10 x 55 8 Self locking nut M10 9 wood screw ø 5x45 10 trigger safety side 11 screw M6x15 pvc ball head 12 spring ø10x30 ...

Page 25: ...x 50 6 laminat for back rest mattressplatform 650x630 23 screw M10 x 55 7 short safety side 24 screw M10 x 60 8 long safety side 25 rubber handle ø12x100 9 bumper wheel 26 spring ø10x30 10 gas cilinder OBKOV NHC C3 130 422 H32P5 500N 27 plastic end cap 30x30 11 plate bumper wheels 28 U frame fo cilinder 140 12 blind rivet ø4x16 29 screw M6x15 pvc ball head 13 plastic end cap ø20 30 trigger safety ...

Page 26: ...26 13 4 Parts list of the mattress and pillow Mattress and pillow art 155040 Art 155040 No item 1 Mattress two section 2 pillow 3 Terminal women 4 Terminal men 5 Drain valve 6 flexible tube ...

Page 27: ...27 Art 155140 ...

Page 28: ...mission of TIMO d o o Ajdovščina FURNITURE FOR HOSPITALS CLINICS NURSING HOMES TIMO d o o Ajdovščina Vipavska cesta 4i 5270 Ajdovščina Slovenija Tel 00386 0 5 36 89 330 Fax 00386 0 5 36 63 805 e mail info timo si Art Izdelka 155040 in 155140 date 25 oktober 2013 rev 00 ...

Reviews: