10 Technical specifications
60
IFU, MIRUS Controller, en-GB, N.00
Specification
Device
Agent supply
Internal reservoir
Maximum capacity = 270 mL
Reserve capacity = 20 mL
Status information
On screen
Filling system
Proprietary filling system, agent specific
according to ISO 5360
Filling capacity
Maximum = 190 mL
Maximum overfill protection = 250 mL
Electromagnetic compatibility
(according to EN 60601-1-2)
Test parameters and limit values can be obtained
from the manufacturer if required.
Display structure
Screen
Touchscreen, 5.7", confirm button
Menu language
English, German, French, Italian, Spanish
Applied Standards
EN 1041
EN 60601-1
EN 60601-1-2
EN 60601-1-8
EN ISO 5360
EN ISO 8835-4
EN ISO 14971
EN ISO 15223-1
EN ISO 21647
EN ISO 80601-2-12
IEC 60529
IEC 62304
IEC 62366-1
Units
CO
2
Pressure (Paw, PEEP)
VA concentration
mmHg, %, kPa
mbar for German, cmH
2
O for all other languages
Vol%