10 Technical specifications
IFU, MIRUS Controller, en-GB, N.00
59
10
Technical specifications
10.1 General specifications
Specification
Device
Physical dimensions (L x H x D)
325 x 195 x 210 mm (12.8 x 7.6 x 8.2 in.)
Weight
with filled reservoir
9.0 kg (15.4 lb)
HW Options
MC-MC-ISO
MC-MC-SEVO
MC-MC-DES
For VA Isoflurane
For VA Sevoflurane
For VA Desflurane
Environmental conditions
Operation (filled reservoir):
Temperature range for MC ISO, SEVO
Temperature range for MC DES
Atm. pressure range
Equivalent altitude (above sea level)
Humidity range
+10 to +40°C
+10 to +30°C
700 to 1,060 hPa
3,000 to 0 m (9,840 to 0 ft)
10 to 90% relative, none condensing
Environmental conditions
Storage with filled or empty reservoir:
Temperature range
Atm. pressure range
Equivalent altitude (above sea level)
Humidity range
-20 to +50°C
500 to 1,060 hPa
5,500 to 0 m (18,050 to 0 ft)
10 to 90% relative, none condensing
Environmental conditions
Transport (empty reservoir):
Temperature range
Atm. pressure range
Equivalent altitude (above sea level)
Humidity range
-20 to +70°C
500 to 1,060 hPa
5,500 to 0 m (18,050 to 0 ft)
10 to 90% relative, none condensing
Noise level
Patient/device alarms at max. setting
“Panic” alarms (fail safe, mains supply
lost, microcontroller watchdog error)
≤ 49 dB (A)
> 66 dB (A) (tone sequence) at max. setting
> 63 dB(A) (permanent tone)
Classifications
CE class according to 93/42/EEC
Protection class acc. to EN 60601-1
Protection type acc. to EN 60601-1
IP Code
IIb
I
B
IP20
GMDN codes
MC-MC-ISO
MC-MC-SEVO
MC-MC-DES
36890
36980
36979
UMDNS code MIRUS Controller
10-144
Electrical supply
Nominal voltage
Frequency
Power consumption
Grounding
Internal back-up
100 to 230 V
AC
± 10%
50 to 60 Hz ± 5%
< 75 VA
Standard ground stud
Built in backup battery, backup time: 15 min.