3 Preparation
22
IFU, MIRUS Controller, en-GB, N.00
3.5
Filling during start-up sequence
If liquid volume level in reservoir is less than 75 mL, MIRUS Controller will require filling prior
to the system test. Wait for a request to fill (message on screen) and follow the instructions.
Note:
For further information about the bottle adapter refer to the manufacturer's instructions.
CAUTION
Only the suitable bottle adapter will match with MIRUS Controller.
CAUTION
Don’t force an unsuitable adapter into the fill port.
CAUTION
Double check bottle adapter colour, fill port collar colour and agent
bottle colour to match dedicated volatile anaesthetic agent prior to
filling.
WARNING
Do not press the fill port valve manually down at any time. This will
result in pressure and vapour release.
WARNING
Slowly remove bottle from fill port to reduce vapour pressure to avoid
ambient pollution.
WARNING
Avoid spilling volatile anaesthetic agent during filling process to avoid
ambient pollution.
Perform
CAUTION
Do not try to open the
flap manually, it could
be seriously damaged.
Press
Fill port flap
button to open fill port
flap or
press Turn off button to shut down device.