6 Faults
If faults cannot be repaired using the table below, stop using the device so as to avoid major
damage.
Fault
Cause
Repair
Power supply LED not
illuminated.
Device not connected to
mains adapter.
Connect device to the mains adapter.
Mains adapter not
connected to the power
supply.
Connect mains adapter to the power
supply.
Mains adapter or power
cable is defective.
Replace mains adapter and power cable.
Data communication
LED not illuminated.
Data communication
cannot be established
because the device is not
connected to the MIRUS
Controller or the patient
monitor.
Connect device to the MIRUS Controller.
Connect device to the EC10 interface on
the patient monitor or the IntelliBridge
EC40/80 hub of the information centre
via the IntelliBridge EC5 ID module.
Data communication
cannot be established
because the MIRUS
Controller is not
connected to the mains
power supply.
Connect the MIRUS Controller to the
mains power supply.
Data communication
LED is flashing.
There is a problem with
the data communication.
Ensure that the USB connectors of the
two connecting cables are correctly
inserted.
No data is displayed on
the patient monitor.
Faults or errors in the
patient monitor.
Refer to the instructions for the patient
monitor.