2 Safety information
10
ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
2.1.7
Delays in transmission
Signals of measurements and alarms from the MIRUS Controller may be delayed before being
transferred to a patient monitor. This delay time may vary depending on the conditions in which the
Data Bridge is working. For this reason, specified delay times cannot be provided.
Only use the MIRUS Controller as the primary source for alarms and for viewing
measurements.
Refer to the manufacturer's instructions regarding patient monitor delay times.
2.1.8
Deviations in transmission
The data transferred to a patient monitor may be displayed incompletely or incorrectly on the
monitor.
Alarms
The various alarm priorities may be displayed differently on the patient monitor than on the MIRUS
Controller connected via the Data Bridge.
Check which alarms are assigned to which alarm priorities.
Only use the MIRUS Controller alarm system as the primary alarm source.
Measurements
In certain cases, there may be differences between the measurements displayed on the patient
monitor and the measurements displayed on the MIRUS Controller connected via the Data Bridge.
Only use the MIRUS Controller as the primary source for viewing measurements.
2.1.9
Modifications to the device
Modifications to this device may lead to malfunctions and unforeseeable hazards, possibly causing
personal injury or damage to property.
Do not modify this device.
2.1.10 Accessories
The Data Bridge has been tested for electromagnetic compatibility with the accessories on the
accessories list. The Data Bridge's electromagnetic emissions may be increased, and the
interference immunity may be reduced, if accessories not included on the accessories list are used.
Moreover, the proper functioning of the Data Bridge may be affected, possibly resulting in personal
injury or damage to property.
Only use the accessories specified by the manufacturer.
Other accessories may only be used if they do not affect the electromagnetic compatibility.