background image

TARE function 

1.  Turn on scale as described above 
2. 

Place the “tare item” on the platform 

3. 

Press “TARE” and wait until “0.0” (or “0”) is displayed 
 
 

CALIBRATION 

1. 

Turn on the scale by pressing ON/OFF key, press “MODE” key quickly and 
hold it until LC

D displays “CAL”. 

2.  Press the MODE key again, the 

LCD flashes “CAL” and ”100.0” (subject to 

the  capacity  of  your  scale)  for  three  times,  now  add  corresponding  cal 
weight on the scale, wait 3-

4 seconds, LCD displays “PASS” and back to 

normal weighing mode.   

3.  Take away the calibration weight and press ON/OFF key to turn off scale, 

the calibration is finished. 

 
 

Care Instructions 

 

Avoiding  any  exposure  to  extreme  heat  or  cold,  your  scale  works  better 
when  operated  at  normal  room  temperature.  Always  allow  the  scale  to 
acclimate to normal room temperature for at least one hour before use. 

 

Keep your scale in a clean environment. Dust, dirt, moisture, vibration, air 
currents and a close distance to other electronic equipment can all cause 
an adverse effect on the reliability and accuracy of your scale. 

 

Handle  with  care.  Gently  apply  all  items  to  be  weighed  onto  tray  top. 
Although  this  scale  is  designed  to  be  quite  durable,  try  to  avoid  rough 
treatment as this may permanently damage the internal sensor. 

 

Avoid  shaking,  dropping  or  otherwise  shocking  the  scale.  This  is  a 
precision instrument and must be handled with extreme care. 