29
5 Maintenance and Service
Danger of possibly contaminated parts!
Contaminated parts can cause personal injuries. Inform yourself
regarding possible contamination before you start working. Be su-
re to follow the relevant instructions and take care of necessary
protective measures.
The unit requires no maintenance. External dirt and soiling can be removed by a
damp cloth.
Should a defect or damage occur on your VSP/VCP, please return the instrument
for repair and enclose a declaration of contamination (see next page).
The unit is not prepared for customer repair!
Malfunction of the unit which is caused by contamination or wear
and tear is not covered by warranty.
Error Messages and Malfunction
Problem
Possible Cause
Correction
output signal 1.4 V
pressure under range
pressure is below range
limit
output signal < 0.5 V
defective electronics or
sensor
send unit for repair
high measurement error
contamination, ageing,
extreme temperature,
maladjustment
readjustment
adjustment not possible
measurement error
exceeds possible range of
readjustment
send unit for repair