Overpressure in the vacuum system > 1 bar
Accidental or unintended opening of clamp elements under stress
can lead to injuries due to parts flying around! Unsecured hose
connections can release, process media thus can leak and possibly
damage your health!
Overpressure in the vacuum system 1.5 to 4 bar
KF flange connections with elastomer sealings cannot withstand
such pressures. Process media thus can leak and possibly damage
your health!
3.3 Electrical Connection
3.3.1 Operation with Thyracont Display Unit
For operation of the transducer with a Thyracont display and control unit a suitable
measurement cable must be used (see accessories).
Do not connect or disconnect the transducer when the cable is on
circuit!
Connect the cables plug to the transducer and secure it with the screw. Connect
the other end of the cable to the display unit and secure the plug. Only now connect
your display unit to mains power or switch it on respectively.