7-
28
DSPA
n
(Dry Bulb Span Coefficient)
DSPA=number
is used during factory calibration, and temporarily changes the
SPAN coefficient of the Dry Bulb temperature formula. The
number
stored is valid until power is removed, the
RESET=C
command is used, the
PROB
e command is used, the probe
number is set or changed from the keypad, or a cold reset is
executed from the keypad. The power-up default value is
determined during factory calibration and permanently stored in
non-volatile memory. Refer to the
REST
ore and
SAVE
commands to alter non-volatile memory.
DSPA?
returns the SPAN coefficient of the Dry Bulb temperature
formula of the currently selected probe.
IBM BASICA Example (Refer to
DLIN
command).
TERMINAL Example (Refer to
DLIN
command).
IEEE 488 Example (Refer to
DLIN
command).
________________________________________________________________________
RELATED COMMANDS:
DLIN
earity
DZER
o
PROBE
REST
ore
SAVE