7-
66
WSPA
n
(Wet Bulb Span Coefficient)
WSPA=number
is used during factory calibration, and temporarily changes the
SPAN coefficient of the Wet Bulb temperature formula. The
number
stored is valid until power is removed, the
RESET=C
command is used, the
PROB=n
command is used, the probe
number is set or changed from the keypad, or a cold reset is
executed from the keypad. The power-up default value is
determined during factory calibration and stored in non-volatile
memory. Refer to
REST
ore and
SAVE
commands to alter
non-volatile memory.
WSPA?
returns the SPAN coefficient of the Wet Bulb temperature
formula of the currently selected probe.
IBM BASICA Example (Refer to
WLIN
command).
TERMINAL Example (Refer to
WLIN
command).
IEEE 488 Example (Refer to
WLIN
command).
________________________________________________________________________
RELATED COMMANDS:
WLIN
earity
WZER
o
PROB
e
REST
ore
SAVE