GPX3800 and GPX3850 Glass Processors
Chapter 7: Maintenence
Page 68
TTN047534-D02
7.8. Measuring the Argon Flow Rates
The argon gas flow rate settings should be adjusted when switching between different fiber diameters.
The flow rate settings are set in the splice file.
To check and/or calibrate the argon flow rates:
Ensure the system is on and the regulator is set to9 – 12 psi.
Remove the filament assembly according to the directions given in Section 7.4: Filament Replacement
Procedure.
Insert the flow meter tube into the argon port as shown in Figure 56. Place the flow meter vertically on a
solid surface for the most accurate reading.
Select
Argon Flow Rate
s from the menu bar. In the popup window, set the desired flow rate. Initiate the
purge using the Purge tool bar item. Wait 5-10 seconds for the flow meter reading to stabilize. Verify that
the flow meter measures the set flow value. Reset the purge flow rate to 2 l/min on completion of the test.
Figure 56
Image of Argon Gas Port (Left). Flow Meter in Vertical Position (Right)
7.9. RMA Process
Contact Thorlabs for information on requesting an RMA for a purchased GPX3800 or GPX3850 glass processor.
ATTENTION
When returning the product to Thorlabs, please use the original packaging and pink anti-static bag
that the unit came in.
The illustrations below show the correct way to re-pack your glass processor workstation. Please ensure that the
bottom of the box is taped securely before re-packing the unit. If returning a malfunctioning unit for repair, we ask
that you return the accessories with the unit in case one of the accessories happens to be the source of the
problems.
RMA Packing Process
!
!