GPX3800 and GPX3850 Glass Processors
Chapter 7: Maintenence
Rev B, July 24, 2018
Page 63
7.
Remove the 2 screws that secure the filament to the furnace body using a 5/64" hex key or balldriver
(See Figure 51).
Figure 51
Remove filament body screws using a 5/64" hex key or balldriver.
8.
Remove the filament body and replace with a new one. Be sure to have the filament “V” number
inscription facing toward the front (user side) of the unit.
9.
Fit and secure the 2 screws that fasten the filament body to the furnace. Do not over tighten (use a Set
the torque of 6 lb-in).
Figure 52
Fit and resecure the filament body to the furnace.
10.
Refit the stainless steel cover plates and secure the two screws at a torque of 6 lb-in.
Figure 53
Secure the stainless steel cover plates on the filament assembly