![THORLABS VEG200 Series User Manual Download Page 30](http://html1.mh-extra.com/html/thorlabs/veg200-series/veg200-series_user-manual_1111028030.webp)
SS-OCT System Base Unit
Chapter 4: System Operation
Page 26
MTN013142-D02
4.1.2. Starting the Software
1) Start the ThorImageOCT software (see Software User’s Manual for details). The system will be switched
on via remote control from the computer. Wait for about 10 seconds until the PC has recognized the
hardware. The green “Sys OK” LED should turn on indicating that the system is ready for use as shown
in Figure 30. Furthermore, the ring-light illumination on the scanner should be turned on.
2) The laser source will be enabled with a short delay once the ThorImageOCT software was started.
Verify that the green “Laser ON” LED is turned on as shown in Figure 30. In case something is wrong
with the laser source or the system is not connected correctly, the software will show the respective
error message.
Figure 30
Software Start: Indicator LEDs
3) In case the interlock pin was removed, a red “Interlock” LED will be turned on as shown in Figure 31.
You will not be able to turn on the laser source. To enable the laser source again, push the interlock
pin into the respective port and depower the base unit (the laser module requires a power cycle for a
reset). The base unit can be powered up again. The “Interlock” LED should be turned off now and the
system is ready for use.
Figure 31
Interlock Error