Stabilized Benchtop IR Light Source
Chapter 6: Maintenance
Rev B, August 4, 2021
Page 15
Flip the orange levers that lock the two yellow wires to their vertical position and pull out the two white cables from
their sockets.
.
Figure 6.4 Disconnect the two white wires.
Then pull the Globar out.
Figure 6.5 Remove the Globar.
To install a new Globar module, insert the new module. Tighten the two locking screws with the included 2 mm
hex key and connect the two yellow wires. Then reinstall the side plate and tighten the four cap screws on the
corners.