6.3 ITC100D Adjustment
The ITC100D is factory adjusted. However, in order to ensure the best display accuracy, it is re-
commended to re-adjust it from time to time.
The adjustment is carried out at a single set point, using the
adjustment potentiometers P1 to
P1 - Adjust I
PD
Inject a known current from an external source into the photo current input (pins
). The current should be about 90% of full scale.
Use P1 to adjust the display reading to the input current value.
P2 - Adjust I
LD
Set the laser controller to constant current mode. Adjust the laser current limit to maximum cur-
rent and the laser current setting to about 90% of maximum current. Connect an ampere meter
to the laser output pins.
Use P2 to adjust the ITC100D reading to the measured output current.
P3 - Adjust I
LD,LIM
Set the laser controller to constant current mode. Adjust the laser current limit to about 90% of
maximum current and the laser current setting to maximum current. Connect an ampere meter
to the laser output pins
Use P3 to adjust the ITC100D reading to the measured output current.
P5 - Adjust T
ACT
Set the temperature sensor to a well defined temperature and adjust the ITC100D reading to
that temperature using P5.
P4 - Adjust T
SET
Adjust the set temperature to about 90% of the full scale value. Turn on the temperature control
and wait until temperature has settled. Read the actual temperature T
ACT
, switch to T
SET
and
bring the reading in accordance with T
ACT
using the potentiometer P4.
Note
T
ACT
must be calibrated first!
P6 - Adjust I
TEC,LIM
Connect an ampere meter to the TEC output pins. Adjust the TEC current limit to about 90% of
full scale. Adjust the temperature set point to a significant higher value and turn on temperature
control.
Adjust the I
TEC,LIM
reading to the ampere meter value using P6.
Summary of Contents for ITC1 Series
Page 1: ...OEM Laser Diode Controller ITC1xx Operation Manual 2018 ...
Page 2: ...Copyright 2018 Thorlabs GmbH Version Date Item No 3 1 10 Jul 2018 M0009 510 501 ...
Page 9: ... 2018 Thorlabs GmbH 2 Getting Started 7 2 3 Operating Elements ...
Page 38: ... 2018 Thorlabs GmbH 36 ITC1xx 8 2 Dimensions ITC1xx All dimensions in mm ...
Page 39: ... 2018 Thorlabs GmbH 8 Appendix 37 8 3 Dimensions ITC100D All dimensions in mm ...
Page 40: ... 2018 Thorlabs GmbH 38 ITC1xx 8 4 Dimensions ITC100F All dimensions in mm ...
Page 46: ...www thorlabs com ...