Page 28
18728-D01
Single- and Dual-Axis Scanning Galvo Systems for Small Beam Diameters
Appendix C Reasons For Image Distortion
The deflection of a laser beam with a two-mirror system results in three effects:
(1) The arrangement of the mirrors leads to a certain distortion of the image field –
see Fig. C.1 below.
Fig. C.1 Field Distortion in a Two-way Mirror Deflection System
This distortion arises from the fact that the distance between mirror 1 and the image
field depends on the size of the mechanical scan angles of mirror 1 and mirror 2. A
larger scan angle leads to a longer distance.
(2) The distance in the image field is not proportional to the scan angle itself, but to
the tangent of the scan angle. Therefore, the marking speed of the laser focus in the
image field is not proportional to the angular velocity of the corresponding scanner.
(3) If an ordinary lens is used for focusing the laser beam, the focus lies on a sphere.
In a flat image field, a varying spot size results.
Beam In
Scanner 1
Scanner 2
Ø2
Ø1
Summary of Contents for GVS001
Page 33: ...Rev 22 Feb 2019 Page 31 Chapter 5 Troubleshooting D 2 CE Certificates ...
Page 34: ...Page 32 18728 D01 Single and Dual Axis Scanning Galvo Systems for Small Beam Diameters ...
Page 35: ...Rev 22 Feb 2019 Page 33 Chapter 5 Troubleshooting ...
Page 36: ...Page 34 18728 D01 Single and Dual Axis Scanning Galvo Systems for Small Beam Diameters ...
Page 37: ...Rev 22 Feb 2019 Page 35 Chapter 5 Troubleshooting ...
Page 38: ...Page 36 18728 D01 Single and Dual Axis Scanning Galvo Systems for Small Beam Diameters ...
Page 39: ...Rev 22 Feb 2019 Page 37 Chapter 5 Troubleshooting ...
Page 41: ...www thorlabs com ...