Single- and Dual-Axis Scanning Galvo Systems for Small Beam Diameters
4.4 Recommended Scanning Angles
The ideal scanning angle is dependent upon a number of conditions. Firstly, the larger
the diameter of the input laser beam, the smaller the achievable scanning angle.
Secondly, the applied input voltage causes the laser beam to move away from the
center of the mirrors. The larger the input voltage then the greater the movement from
the center, as shown below.
Lastly, on dual-axis systems, there is an offset alignment between the X and Y axis
mirrors that also limits the scan angle.
The table below gives recommended scanning angles for various beam diameters.
GVS001, GVS101, GVS201 and GVS301,
Input Beam Diameter
Mechanical Scan Angle
Optical Scan Angle
4 mm and less
± 12.5°
± 25°
5 mm
+ 10°, -12.5°
+ 20°, -25°
GVS002, GVS102, GVS202 and GVS302,
Input Beam
Diameter
Mechanical
Scan Angle X
Optical
Scan Angle X
Mechanical
Scan Angle Y
Optical
Scan Angle Y
1 mm
+ 11.0°, - 12.5°
+ 22.0°, - 25°
± 12.5°
± 25°
2 mm
+ 10°, - 11.5°
+ 20°, - 23°
± 12.5°
± 25°
3 mm
+ 9.5°, - 10°
+ 19°, - 20°
± 12.5°
± 25°
4 mm
± 8.5°
± 17°
± 12.5°
± 25°
5 mm
± 8°
± 16°
+ 12.5°, -3°
+ 25, -6°
Summary of Contents for GVS001
Page 33: ...Rev 22 Feb 2019 Page 31 Chapter 5 Troubleshooting D 2 CE Certificates ...
Page 34: ...Page 32 18728 D01 Single and Dual Axis Scanning Galvo Systems for Small Beam Diameters ...
Page 35: ...Rev 22 Feb 2019 Page 33 Chapter 5 Troubleshooting ...
Page 36: ...Page 34 18728 D01 Single and Dual Axis Scanning Galvo Systems for Small Beam Diameters ...
Page 37: ...Rev 22 Feb 2019 Page 35 Chapter 5 Troubleshooting ...
Page 38: ...Page 36 18728 D01 Single and Dual Axis Scanning Galvo Systems for Small Beam Diameters ...
Page 39: ...Rev 22 Feb 2019 Page 37 Chapter 5 Troubleshooting ...
Page 41: ...www thorlabs com ...