background image

Electronic Variable Optical Attenuators

 

Chapter 3: Introduction & Quick Start 

Page 6 

 

TTN095737-D02 

3.4.  Quick Start Guide 

ESD CAUTION 

The components inside these products are sensitive to electrostatic 

discharge (ESD). Take all appropriate precautions, including grounding 

personnel and equipment, before making any connections to the unit. 

In the box, you should find the EVOA and its power supply with a 2.1 mm 
coaxial plug that matches the power input jack on the side of the unit. 

Follow these simple instructions to get started as quickly as possible: 
 

1. 

Make sure the unit is sufficiently charged.

 It is shipped about 50% 

charged. Using the included power supply, it typically takes 5 hours 
to reach a full charge (starting from the “Low Battery” indicator) that 
lasts for hundreds of hours. 

2. 

Connect input and output fiber patch cables (single mode only) 
to the bulkheads.

 Make sure they are clean. We recommend the 

FCC-7020 Fiber Connector Cleaner and the FBC1 Bulkhead Cleaner 
for fiber tips and bulkheads, respectively. 

3. 

Select the desired wavelength with the selector knob.

 The unit is 

factory calibrated for both 1310 nm and 1550 nm. Other wavelengths 
in the 1250 - 1625 nm range can be user calibrated. 

4. 

Turn the rocker power switch ON.

 

5. 

Adjust the attenuation to get the desired output power on the 
digital display.

 

6. 

Press the Lock button to hold the power setting (optional).

 A 

green LED will blink to indicate a successful lock. A red LED will blink 
if the unit cannot find a lock. 

7. 

Release the Lock button to make any further adjustments.

 

Summary of Contents for EVOA1550A

Page 1: ...EVOA1550F EVOA1550A Electronic Variable Optical Attenuators User Guide...

Page 2: ...nstructions Details 7 4 1 Overview 7 4 2 Powering Charging the Unit 7 4 3 Fiber Connections 8 4 4 Wavelength Selection Calibration 8 4 5 Attenuation Control 9 4 6 Digital Output Power Display 9 4 7 Mo...

Page 3: ...Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Off Supply In Position...

Page 4: ...rdance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation Thorlabs is not responsi...

Page 5: ...d accept either FC PC connectors Item EVOA1550F or FC APC connectors Item EVOA1550A The wavelength selector switch accesses calibrated settings for 1310 nm and 1550 nm and also provides a position whi...

Page 6: ...3 Fiber Interface Utility Panels Rocker Power Switch 2 1 mm Charging Jack for Included Power Supply USB Mini B Charging Jack Charging Indicators Green Fully Charged Yellow Charging SMA Modulation Inpu...

Page 7: ...ally takes 5 hours to reach a full charge starting from the Low Battery indicator that lasts for hundreds of hours 2 Connect input and output fiber patch cables single mode only to the bulkheads Make...

Page 8: ...controlled by the manual power adjustment knob the Power Lock control loop or the SMA modulation input When the power is switched off the EVOA defaults to a mid range attenuation whose value is indep...

Page 9: ...at uses virtually no current as the EVOA is an ultra low current device This feature does not drain the battery meaningfully faster than the battery itself would lose charge otherwise 4 3 Fiber Connec...

Page 10: ...panel see Figure 1 This knob provides the full range of attenuation Rotating it clockwise increases the output power while rotating it counterclockwise decreases the output power The knob has ten full...

Page 11: ...wer setpoint This circuit has a maximum bandwidth of 1 Hz Note that pushing the button may cause a slight offset in the output power which is due to the resolution of the sample and hold circuit If th...

Page 12: ...t thorlabs com Do not disassemble the unit There are no user serviceable components inside The end faces of any fiber patch cables being connected to the EVOA should be cleaned every time a connection...

Page 13: ...V to 4 9 V Absolute Maximum 5 0 V Input Impedance High Z Bandwidth DC to 1 kHz Typ Connector Female SMA Monitor Output Conversion Gain 20 mV mW Output Impedance High Z Bandwidth DC to 1 kHz Min DC to...

Page 14: ...Electronic Variable Optical Attenuators Chapter 6 Specifications Rev B September 5 2017 Page 13 6 2 Mechanical Drawings 3 1 mm 0 12 55 2 mm 2 17 166 9 mm 6 57 161 0 mm 6 34 87 0 mm 3 43 93 2 mm 3 67...

Page 15: ...other Thorlabs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassem...

Page 16: ...Electronic Variable Optical Attenuators Chapter 8 Certificate of Compliance Rev B September 5 2017 Page 15 Chapter 8 Certificate of Compliance...

Page 17: ...anada and South America Thorlabs Inc sales thorlabs com le techsupport thorlabs com Europe Thorlabs GmbH europe thorlabs com France Thorlabs SAS SA sales fr thorlabs com Japan Thorlabs Japan Inc sales...

Page 18: ...www thorlabs com...

Reviews: