© 2020 Thorlabs GmbH
4 Write Your Own Application
43
4.3.2 SCPI Command Reference
SYSTem subsystem commands
Command
Description
SCPI
SYSTem
Path to SYSTem subsystem
Vol.2 §21
:BEEPer
Vol.2 §21.2
[:IMMediate]
Issues an audible signal
Vol.2 §21.2.2
:STATe {ON|1|OFF|0}
Activates/deactivates the beeper
Vol.2 §21.2.3
:STATe?
Returns the state of the beeper
Vol.2 §21.2.3
:VOLume <value>
Sets the beeper volume in a range from 0.0 to 1.0
Vol.2 §21.2.5
:VOLume?
Returns the beeper volume
Vol.2 §21.2.5
:ERRor
Vol.2 §21.8
[:NEXT]?
Returns the latest error code and message
Vol.2 §21.8.8
:VERSion?
Returns level of SCPI standard (1999.0)
Vol.2 §21.21
:TERMinal{[1]|2}
Information about the LED heads connected to the
LED terminals LED1 or LED2
[:HTYPe]?
LED head type identification query. The query re-
sponse is in the format
"<vendor_-
name_string>,<led_head_model_-
name_string>,<led_head_seri-
al_no_string>,<fw_version_ma-
jor_num>,<fw_ver-
sion_minor_num>,<fw_version_sub-
minor_num>"
.
The firmware version indicates the head’s onboard
memory mapping version. A major version number
-1 indicates a custom head without onboard head
info memory. A major version number – 2 indicates
that there is no head connected on the terminal
Sample of a LED with onboard head info memory:
Thorlabs,M365F1,M00123456,1.0.0
Sample of a LED without onboard head info
memory:
Thorlabs,custom,n/a,-1.0.0
Sample of no LED head connected:
Thorlabs,no head,no head,-2.0.0
:HEAD
:TEMPerature
[:COUNt]?
Get the number of temperature sensor available in
the LED head currently connected to the specified
terminal.
:LABel? [{0|<sensidx>}]
Query LED head temperature sensor label.
This is a query with parameter. The optional nu-
meric parameter is the index of the LED head tem-
perature sensor to query. Parameter value range is
0 <= sensidx < senscnt with default parameter val-
ue is 0.
:VOLTage?
Query maximum LED forward voltage specified by
the head’s onboard info memory
:CURRent?
Query maximum LED forward current specified by
the head’s onboard info memory
:SPECtrum?
Query the spectrum information of the LED head.
Positive values describe the nominal LED center
wavelength in nm. Negative values describe the
color temperature in K. The value 0.0 describes "in-
formation not available"
Summary of Contents for DC2200
Page 1: ...High Power LED Controller DC2200 Operation Manual 2020 ...
Page 2: ...Copyright 2020 Thorlabs GmbH Version Date 1 6 23 Sep 2020 ...
Page 31: ... 2020 Thorlabs GmbH 3 Operating Instruction 29 Click Next and then Install to continue ...
Page 70: ... 2020 Thorlabs GmbH 68 DC2200 6 2 Dimensions ...
Page 74: ... 2020 Thorlabs GmbH 72 DC2200 6 7 Certifications and Compliances ...
Page 78: ...www thorlabs com ...