© 2020 Thorlabs GmbH
20
DC2200
Note
The frequency range of the modulating TTL signal is measured under the following definitions
and conditions:
·
Resistive load
·
The "Maximum Modulation Frequency"
f
max
is defined as the frequency, when the sum of
rise time
t
rise
and fall time
t
fall
equals 10 % of the TTL signal period:
·
Rise and fall time are defined as the time between 10% and 90% of the current at TTL
HIGH level
Current Range
Measurement Current
Load
Max. Modulation Frequency
Terminal LED1
0.0 - 4.0 A
500 mA
1.0
W
23 kHz
4.0 - 10.0 A
1000 mA
0.5
W
18 kHz
Terminal LED1
0.0 - 2.0 A
250 mA
2.0
W
27 kHz
3.3.8 Info
Tap the
button to display information about the DC2200 hardware and firmware: