36
HA0278T Rev H Dec 2016
Chapter 5
Jog
- used to increment or decrement the motor position. When the button is clicked,
the motor is driven in the selected direction at the jog velocity one step per click. The
step size and jog velocity parameters are set in the 'Settings' panel (see Section 5.3.).
Travel
- displays the range of travel (in millimeters or degrees) of the motor.
Moving
- lit when the motor is in motion.
Enable
- applies power to the motor. With the motor enabled, the associated Channel
LED on the front panel is lit.
Digital display
- shows the position (in millimetres or degrees) of the motor. The
motor must be ' Homed' before the d isplay will show a valid position value, (i.e. the
displayed position is relative to a physical datum, the limit switch).
Home
- sends the motor to its 'Home' position - see Section F.2.2. The LED in the
button is lit while the motor is homing.
Homed
- lit when the motor has previously been 'Homed' (since power up).
Stop
- halts the movement of the motor.
Limit switches
- the LEDs are lit when the associated limit switch has been activated
- see Section F.2.3. for further details on limit switches.
Settings display
- shows the following user specified settings:
Driver -
the type of control unit associated with the specified channel.
Stage
- the stage type and axis associated with the specified channel.
Calib File
- the calibration file associated with the specified channel.
See the APTConfig utility helpfile for more details on assigning and using calibration
files.
Min/Max V
- the minimum velocity at which a move is initiated, and the maximum
velocity at which the move is performed. Values are displayed in real world units (mm/
s or degrees/s), and can be set via the 'Settings' panel (see Section 5.3.).
Accn
- the rate at which the velocity climbs to, and slows from, maximum velocity,
displayed in real world units (mm/s/s or degrees/s/s). The acceleration can be set via
the 'Settings' panel (see Section 5.3.) and i s used in conjunction with the Min/Max
velocity settings to determine the velocity profile of a motor move. See Appendix F
Section F.1.3. for more information on velocity profiles.
Jog Step Size
- the size of step (in mm or degrees) taken when the jog signal is
initiated. The step size can be set e ither via the Settings panel or by calling the
SetJogStepSize method.
Settings button
- Displays the 'Settings' panel, which allows the motor drive’s
operating parameters to be entered - see Section 5.3.
Ident
- when this button is pressed, the LED (on the front panel of the unit) associated
with the selected channel will flash for a short period.
Active
- lit when the unit is operating normally and no error condition exists.
Error
- lit when a fault condition occurs.