5.3 Electrical Mounting for Precipitation Sensor with Plug
Connection
For: 5.4103.20.741, see circuit diagram in chapter 5
5.3.1 Plug Mounting
Applies only to instruments with connection „plug“.
Coupling socket, Typ:Binder, Serial 423, EMC with cable clamp
Cable connection: without cable shield
1. Stringing parts on cable acc. to plan given above.
2. Stripping cable sheath 20 mm
3. Cutting uncovered shield 20 mm
4. Stripping wire 5mm.
5. Soldering wire to the insert
6. Positioning shield in cable clamp.
7. Screwing-on cable clamp.
8. Assembling remaining parts acc. to upper plan.
9. Tightening pull-relief of cable by screw-wrench (SW16 und
17).
Cable clamp
Cable sheath
Wire
6 Taking into Operation
After the electrical connection has been established, and the case has been screwed, the
operating voltage can be switched on. The dimension of the measurement value output is
reset when the supply voltage is switched on.
7 Maintenance
The device is maintenance free.
Cleaning
:
The pollution and the pollution level is dependent on the location. Therefore, we recommend
the unit be checked at appropriate intervals and cleaned if necessary.
For the cleaning should use a damp cloth without chemical cleaning agents are used.