   1

Battery Pack

© THI GmbH 2020/05

REF 20602 | Rev-02

BATTERY PACK

R

X

 ONLY

INSTRUCTIONS FOR USE 

TECHNICAL SPECIFICATION

MODEL

REF 60401

Description

Rechargeable Battery Pack

European Conformity

CE

Cell chemistry

Lithium Ion

Voltage (nominal)

14.8 V (DC)

Capacity (rated)

2.950 mAh (43.6 Wh)

Cycle life characteristics

> 80 % of rated capacity after 400 duty cycles

Ingress Protection (IP)

IPX0

Mode of operation

Continuous

Charging time

4 hours

Battery life

> 5 hours  (with REF 80100 

ViVi

®

 Helmet)

> 3 1/2 hours  (with REF 80120 

ViVi

®

 Helmet w/Headlight)

> 6 hours (with REF 70100, 70110 

   OXYGEN 

Aerolyte Helmets)

> 4 ½ hours (with REF 70120, 70130 
  

OXYGEN 

Aerolyte LED Helmets)

Mass

250 g  (8.8 oz)

LIMITATIONS OF USE

WARNINGS:  

• Upon delivery and before each use, inspect the battery pack to check for any 

eventual damage. DO NOT drain the battery if it seems damaged. DO NOT use 
a battery that emits odors, generates heat, or shows discoloration or leakage.

• 

DO NOT expose the battery pack to heat, fi re, or shock.

• DO NOT heat the battery over 80 °C (176 °F).
• Exposure to high temperatures can reduce battery capacity and performance.
• DO NOT expose the battery to sources of heat or sunlight.
• DO NOT use this equipment in the presence of mixtures containing anesthetic 

substances, fl ammable air, oxygen or nitrous oxide.

• Unless otherwise indicated, DO NOT sterilize or immerse the battery pack in 

water, solvents, lubricants or other cleansing chemicals, or even expose them to 
those substances. Keep the battery clean and dry.

• DO NOT short-circuit, crush, open, tear, incinerate, or disassemble the bat-

tery pack. Never store a bulk of batteries in a box or drawer where they could 
short-circuit or be short-circuited by metallic objects. Damage increases the 

risk of chemical burns due to electrolyte spills, fi re or explosion. For replacement 

instructions, contact your local distributor.

• In case of leakage of liquids from the battery, avoid skin contact. In case of con-

tact, wash the aff ected area with plenty of water and consult a doctor. In case of 

ingestion of a cell or whole battery pack, consult your doctor immediately. Keep 
the battery pack out of reach of children.

• 

DO  NOT  use  batteries  other than those  specifi cally  designed for the 

OXYGEN 

and 

ViVi

®

 System. The use of any other battery may cause fi re, explosion, elec-

tric shock, or injury.

• Use the battery pack only for the intended applications.
• When using medical equipment, such as this battery pack, comply with the elec-

tromagnetic compatibility (CEM) procedures provided by your hospital.

RECOMMENDATIONS FOR BATTERY HANDLING

• DO NOT remove the battery from its original packaging until it is used.
• Upon delivery, your battery is in a shipping mode and will not deliver power. 

Before fi rst use, put the battery into the charger to wake it up.

• After long storage periods, a full charging/discharging cycle might be necessary 

to regain full battery capacity.

• Dispose the battery pack as described in the Disposal / Recycling section of 

this manual.

• Rechargeable batteries achieve their optimum performance at room tempera-

ture, at 20 °C ± 5 °C (68 °F ± 10 °F).

• Store the original product documentation for future reference.

CHARGING INSTRUCTIONS

• For detailed instructions, consult the operator manual of the 

ViVi

®

 Helmet and 

OXYGEN

 Helmet.

WARNINGS:  

• ALWAYS use the REF 60500 Battery Charger to charge and store the battery pack. 

60401 BATTERY PACK

INSTRUCTIONS FOR USE - ENGLISH

IMPORTANT INFORMATION

The words 

WARNING

CAUTION

, and 

NOTE

 have special meaning and should be 

reviewed. 

WARNING

WARNING

 highlights a safety-related issue. 

ALWAYS

 comply 

with this information to prevent patient and/or healthcare staff  

injury.

CAUTION

CAUTION

 highlights a product reliability issue. 

ALWAYS

 comply 

with this information to prevent product damage.

NOTE

NOTE

 supplements and/or clarifi es procedural information. 

COMPLIANCE STATEMENTS

This rechargeable Lithium Ion battery pack REF 60401 has been tested and has met 
the requirements of the following applicable regulations:

• IEC 62133:2002
• FCC Title 47 CFR, Part 15, Class B/ IECS-003, IV edition
• ROHS Directive 2002/95/EC
• Batteries Directive 2006/66/EC and subsequent amendments
• EMC Directive 2004/108/EC
• UN Transportation Testing – UN DOT 38.3 (T1-T8) Lithium Batteries

INDICATIONS FOR USE

This battery pack is designed for use with the 

OXYGEN

 and 

ViVi

®

 System. These 

Systems are intended to be worn by surgical personnel to provide a barrier be-
tween the operating environment and the surgical personnel in order to help pro-

tect against contamination and/or exposure of infectious body fl uids and harmful 

microorganisms. 

FOR USE WITH

• Use this battery pack to power:

REF 80100 

ViVi

®

 Helmet, 

REF 80120 

ViVi

®

 Helmet w/ Headlight, 

REF 70100 

OXYGEN 

Aerolyte Helmet XS-XL, 

REF 70110 

OXYGEN 

Aerolyte Helmet M-4XL, 

REF 70120 

OXYGEN 

Aerolyte LED Helmet XS-XL, 

REF 70130 

OXYGEN 

Aerolyte LED Helmet M-4XL.

• Charge this battery pack only with REF 60500 Battery Charger.

SYMBOL DEFINITIONS

These symbols are located on the device and/or labels and are defi ned as follows:

SYMBOL

DEFINITION

Caution, consult accompanying documents

Waste Electrical and Electronic Equipment Directive (WEEE)

LI-ion

Recyclable Product (see section ‘

Disposal/Recycling’

)

Temperature Limitation

Direct Current (DC)

CE Mark

Serial Number

Catalog number

Manufacturer

ENGLISH (EN) - DEUTSCH (DE) - ITALIANO (IT)

Summary of Contents for 60401

Page 1: ...OT remove the battery from its original packaging until it is used Upon delivery your battery is in a shipping mode and will not deliver power Before first use put the battery into the charger to wake it up After long storage periods a full charging discharging cycle might be necessary to regain full battery capacity Dispose the battery pack as described in the Disposal Recycling section of this m...

Page 2: ... wurden zur Verwendung mit dem OXYGEN und ViVi System ent wickelt Diese Systeme werden vom chirurgischen Personal getragen um eine Barriere zwischen dem Operationsfeld und dem chirurgischem Personal zu schaffen und um der Kontamination bzw der Exposition durch infektiöse Körperflüssigkeiten und gefährliche Mikroorganismen vorzubeugen zu helfen VERWENDEN SIE DIESE BATTERIEN ZUR STROMVERSORGUNG VON R...

Page 3: ...r Europäischen EU WEEE Richtlinie 2002 96 EC muss das Produkt getrennt gesammelt werden NICHT als unsortierten Restmüll entsorgen Kontaktieren Sie Ihren lokalen Vertragshändler für Informationen zur Entsorgung Enthält keine schädlichen Substanzen gemäß RoHS Standard HINWEISE Diese Batterie enthält Lithium Am Ende der Nutzungsdauer muss die Batterie durch einen qualifizierten Recycling oder Sonderm...

Page 4: ...aricamento prima che il pacco batteria riacquisti la piena capacità Smaltire il pacco batteria come descritto nella sezione Smaltimento riciclo delle presenti istruzioni d uso Le batterie ricaricabili raggiungono un livello di prestazioni ottimale a tempera tura ambiente ossia a 20 C 5 C 68 F 10 F Conservare con cura la documentazione originale del prodotto per future con sultazioni ISTRUZIONI PER...

Reviews: