EN - DE - IT
1
Battery Pack
© THI GmbH 2020/05
REF 20602 | Rev-02
BATTERY PACK
R
X
ONLY
INSTRUCTIONS FOR USE
TECHNICAL SPECIFICATION
MODEL
REF 60401
Description
Rechargeable Battery Pack
European Conformity
CE
Cell chemistry
Lithium Ion
Voltage (nominal)
14.8 V (DC)
Capacity (rated)
2.950 mAh (43.6 Wh)
Cycle life characteristics
> 80 % of rated capacity after 400 duty cycles
Ingress Protection (IP)
IPX0
Mode of operation
Continuous
Charging time
4 hours
Battery life
> 5 hours (with REF 80100
ViVi
®
Helmet)
> 3 1/2 hours (with REF 80120
ViVi
®
Helmet w/Headlight)
> 6 hours (with REF 70100, 70110
OXYGEN
Aerolyte Helmets)
> 4 ½ hours (with REF 70120, 70130
OXYGEN
Aerolyte LED Helmets)
Mass
250 g (8.8 oz)
LIMITATIONS OF USE
WARNINGS:
• Upon delivery and before each use, inspect the battery pack to check for any
eventual damage. DO NOT drain the battery if it seems damaged. DO NOT use
a battery that emits odors, generates heat, or shows discoloration or leakage.
•
DO NOT expose the battery pack to heat, fi re, or shock.
• DO NOT heat the battery over 80 °C (176 °F).
• Exposure to high temperatures can reduce battery capacity and performance.
• DO NOT expose the battery to sources of heat or sunlight.
• DO NOT use this equipment in the presence of mixtures containing anesthetic
substances, fl ammable air, oxygen or nitrous oxide.
• Unless otherwise indicated, DO NOT sterilize or immerse the battery pack in
water, solvents, lubricants or other cleansing chemicals, or even expose them to
those substances. Keep the battery clean and dry.
• DO NOT short-circuit, crush, open, tear, incinerate, or disassemble the bat-
tery pack. Never store a bulk of batteries in a box or drawer where they could
short-circuit or be short-circuited by metallic objects. Damage increases the
risk of chemical burns due to electrolyte spills, fi re or explosion. For replacement
instructions, contact your local distributor.
• In case of leakage of liquids from the battery, avoid skin contact. In case of con-
tact, wash the aff ected area with plenty of water and consult a doctor. In case of
ingestion of a cell or whole battery pack, consult your doctor immediately. Keep
the battery pack out of reach of children.
•
DO NOT use batteries other than those specifi cally designed for the
OXYGEN
and
ViVi
®
System. The use of any other battery may cause fi re, explosion, elec-
tric shock, or injury.
• Use the battery pack only for the intended applications.
• When using medical equipment, such as this battery pack, comply with the elec-
tromagnetic compatibility (CEM) procedures provided by your hospital.
RECOMMENDATIONS FOR BATTERY HANDLING
• DO NOT remove the battery from its original packaging until it is used.
• Upon delivery, your battery is in a shipping mode and will not deliver power.
Before fi rst use, put the battery into the charger to wake it up.
• After long storage periods, a full charging/discharging cycle might be necessary
to regain full battery capacity.
• Dispose the battery pack as described in the Disposal / Recycling section of
this manual.
• Rechargeable batteries achieve their optimum performance at room tempera-
ture, at 20 °C ± 5 °C (68 °F ± 10 °F).
• Store the original product documentation for future reference.
CHARGING INSTRUCTIONS
• For detailed instructions, consult the operator manual of the
ViVi
®
Helmet and
OXYGEN
Helmet.
WARNINGS:
• ALWAYS use the REF 60500 Battery Charger to charge and store the battery pack.
60401 BATTERY PACK
INSTRUCTIONS FOR USE - ENGLISH
IMPORTANT INFORMATION
The words
WARNING
,
CAUTION
, and
NOTE
have special meaning and should be
reviewed.
WARNING
A
WARNING
highlights a safety-related issue.
ALWAYS
comply
with this information to prevent patient and/or healthcare staff
injury.
CAUTION
A
CAUTION
highlights a product reliability issue.
ALWAYS
comply
with this information to prevent product damage.
NOTE
A
NOTE
supplements and/or clarifi es procedural information.
COMPLIANCE STATEMENTS
This rechargeable Lithium Ion battery pack REF 60401 has been tested and has met
the requirements of the following applicable regulations:
• IEC 62133:2002
• FCC Title 47 CFR, Part 15, Class B/ IECS-003, IV edition
• ROHS Directive 2002/95/EC
• Batteries Directive 2006/66/EC and subsequent amendments
• EMC Directive 2004/108/EC
• UN Transportation Testing – UN DOT 38.3 (T1-T8) Lithium Batteries
INDICATIONS FOR USE
This battery pack is designed for use with the
OXYGEN
and
ViVi
®
System. These
Systems are intended to be worn by surgical personnel to provide a barrier be-
tween the operating environment and the surgical personnel in order to help pro-
tect against contamination and/or exposure of infectious body fl uids and harmful
microorganisms.
FOR USE WITH
• Use this battery pack to power:
REF 80100
ViVi
®
Helmet,
REF 80120
ViVi
®
Helmet w/ Headlight,
REF 70100
OXYGEN
Aerolyte Helmet XS-XL,
REF 70110
OXYGEN
Aerolyte Helmet M-4XL,
REF 70120
OXYGEN
Aerolyte LED Helmet XS-XL,
REF 70130
OXYGEN
Aerolyte LED Helmet M-4XL.
• Charge this battery pack only with REF 60500 Battery Charger.
SYMBOL DEFINITIONS
These symbols are located on the device and/or labels and are defi ned as follows:
SYMBOL
DEFINITION
Caution, consult accompanying documents
Waste Electrical and Electronic Equipment Directive (WEEE)
LI-ion
Recyclable Product (see section ‘
Disposal/Recycling’
)
Temperature Limitation
Direct Current (DC)
CE Mark
Serial Number
Catalog number
Manufacturer