3.
Physical Alignment
(1)
(2)
NOTE:
To avoid undue stress and strain on the
cable connection, suitably Tie-wrap them to the
tripod or the support of choice to reduce potential
damage.
1.
Mount the FSD01 camera (1) on a suitable platform (tripod/ceiling mount)
2.
Mount the ThermaRef 35B1 blackbody (2) on a suitable platform (tripod/ceiling
mount) at a distance of 1-3m from the FevIR Scan 2 FSD01 camera.
3.
When viewed from behind the FevIR Scan 2 FSD01 camera, the ThermaRef
35B1 blackbody should be offset to the left or right and not viewed directly in
line with the front of the field of view.
4.
For optimum system performance the ThermaRef 35B1 blackbody should be
located at the same point as that used to
measure subject’s skin temperature.
This point should be outlined on the floor with tape and people instructed to
pause momentarily to confirm no alarm is activated.
5.
Tape or barriers should be used to ensure people cannot walk between the
camera and reference source.
1-3m
FevIR Scan 2
FSD01 camera
ThermaRef 35B1
blackbody