Thermosense HL-800K User Manual Download Page 7

 

HL-800K Infrared Laser Thermometer User Manual 

 

Thermosense Limited 

T: +44 (0)1628 531166 

E: 

[email protected]

 

W: www.thermosense.co.uk 

Page 7 

How to obtain Emissivity? 

In the EMS mode, press and hold the

 

Laser/Backlight button until the “EMS” icon on the left side of 

LCD blinking. At this time, “ε= --” will appear at the upper side of the LCD; IR temperature value will be 
in the middle of LCD; and Type K temperature value is at the lower side of the LCD.  
Contact the Type K probe to the object surface and test the temperature of same point with IR 
measurement. After both values are stable, press UP or DOWN button for confirmation. Then, 
emissivity of the object will be shown at the upper side of the LCD. Press MODE button or trigger to 
enter normal measurement. Note: 

1. 

When the IR value is not corresponding to the TK value, or IR and TK tested the different points, 
no emissivity will be obtainable or a wrong emissivity will be obtained. 

2. 

The temperature of the target should be higher than the environment temperature. Normally, 
100

 is suitable for obtaining a higher accuracy emissivity. 

After obtaining the emissivity, if the difference between IR value (in the middle of LCD) and TK 
value (at the lower side of LCD) is too big, the obtained emissivity will be incorrect. It’s necessary 
to obtain a new emissivity.

 

 

Note: Measurement Considerations 

Holding the meter by its handle, point the IR Sensor toward the object 
whose temperature is to be measured. The meter automatically 
compensates for temperature deviations from ambient temperature. 
Keep in mind that it will take up to 30 minutes to adjust if wide ambient 
temperatures are to be measured followed by high temperature 
measurements, some time (several minutes) is required after the low 
(and before the high) temperature measurements are made. 
This is a result of the cooling process, which must take place for the IR 
sensor.

 

 

BATTERY REPLACEMENT 

1) 

If battery power is not sufficient, the LCD will display         A new 9V battery is required. 

2) 

Open the battery cover, take out the battery and replace with a new 9-Volt battery and close 
the battery cover.    

 
NOTES 

 

How it Works 

Infrared thermometers measure the surface temperature of an object. The unit’s optics sense 
emitted, reflected, and transmitted energy, which is collected and focused onto a detector. The 
unit’s electronics translate the information into a temperature reading, which is display on the unit. 
In units with a laser, the laser is used for aiming purposes only. 

 

Field of View 

Make sure that the target is larger than the unit’s spot size. The smaller the target, the closer you should be 

to it. When accuracy is critical, make sure the target is at least twice as large as the spot size. 

Summary of Contents for HL-800K

Page 1: ...HL 800K Infrared Laser Thermometer 50 C to 800 C 13 1 ratio User Manual...

Page 2: ...s thermosense co uk W www thermosense co uk Page 2 TABLE OF CONTENTS INTRODUCTION 3 FEATURES 3 WIDE APPLICATION RANGE 3 SAFETY 3 DISTANCE SPOT SIZE 4 SPECIFICATIONS 4 FRONT PANEL DESCRIPTION 5 INDICAT...

Page 3: ...urement Type K temperature Measurement Unique flat surface modern housing design Built in laser pointer Automatic Data Hold Automatic Power Off switch Emissivity Digitally adjustable from 0 10 to 1 0...

Page 4: ...F 20 to 200 C 4 to 392 F 1 5 of reading 2 C 3 6 F 200 to 538 C 392 to 1000 F 2 0 of reading 2 C 3 6 F 538 to 800 C 1000 to 1472 F 3 5 of reading 5 C 9 F Response time Less than 1 second Spectral resp...

Page 5: ...DICATOR 1 Data hold 2 Measuring indication 3 Emissivity symbol and value 4 C F symbol 5 Auto obtain Emissivity 6 Lock and laser on symbols 7 High alarm and low alarm symbol 8 Temperature values for th...

Page 6: ...while the unit is locked on the laser and backlight will be turned on if they have been activated When the unit is locked on the backlight and laser will remain on unless it is turned off using the La...

Page 7: ...ty will be incorrect It s necessary to obtain a new emissivity Note Measurement Considerations Holding the meter by its handle point the IR Sensor toward the object whose temperature is to be measured...

Page 8: ...vity of 0 95 pre set in the unit Inaccurate readings will result from measuring shiny or polished metal surfaces To compensate cove the surface to be measured with masking tape or flat black paint All...

Reviews: