HL-800K Infrared Laser Thermometer User Manual
Thermosense Limited
T: +44 (0)1628 531166
E:
W: www.thermosense.co.uk
Page 4
DISTANCE & SPOT SIZE
As the distance (D) from the object increases, the spot size (S) of the area measured by the unit
becomes larger. The relationship between distance and spot size for each unit is listed below. The
focal point for each unit is 914mm (36”). The spot sizes indicate 90% encircled energy.
SPECIFICATIONS
Thermocouple Measurement
Temperature Range
-50 to 1370°C (-58 to 2498°F)
Resolution
-50 to 1370°C
0.1°C
-58 to 1999°F
0.1°F
2000 to 2498°F
1°F
Accuracy
-50 to 1000°C
±1.5% of reading ± 3°C (±°F)
1000 to 1370°C
±1.5% of reading ± (±3.6°F)
IR Measurement
Temperature Range
-50 to 800°C (-58 to 1472°F)
D : S
13:1
Resolution
0.1°C (0.1°F)
Accuracy
-50 to -20°C (-58 to -4°F)
±5°C (±9°F)
-20 to 200°C (-4 to 392°F)
±1.5% of reading ±2°C (±3.6°F)
200 to 538°C (392 to 1000°F)
±2.0% of reading ±2°C (±3.6°F)
538 to 800°C (1000 to 1472°F)
±3.5% of reading ±5°C (±9°F)
Response time
Less than 1 second
Spectral response
8~14um
Emissivity
Digitally adjustable from 0.10 to 1.0
Over range indication
LCD will show “-0L”,”0L”
Polarity
Automatic (no indication for positive); Minus (-) sign for negative
Diode laser
output <1mW, Wavelength 630~670nm, Class 2(II) laser product
Operating temp.
0 to 50
℃
(32 to 122
℉
)
Storage temp.
–20 to 60
℃
(-4 to 140
℉
)
Relative humidity
10%~90%RH operating, <80%RH storage
Power supply
9V battery, NEDA 1604A or IEC 6LR61, or equivalent
Weight
290g (10.2 oz.)
Size
100 x 56 x 230mm (3.9 x 2.2 x 9.0”)
Safety
“CE” Comply with EMC
Note:
Accuracy:
Given at 18 to 28°C (64 to 82°F), less than 80 % RH.
Field of View:
Make sure that the target is larger than the
unit’s spot size. The smaller the target,
the closer you should be to it. When accuracy is critical, make sure the target is at least twice as
large as the spot size.
Fig: 1