b. RH Calibration
Equipment required: Humidity sensing instrument of known
accuracy.
1. Place the reference humidity instrument in the center of
the chamber.
2. Allow the incubator to run at the desired humidity level
for at least an hour after it has stabilized.
3. Record the humidity level shown by the reference
instrument and the humidity level displayed on the RH
Control module.
4. Turn the Program/Run key switch to Program.
5. Turn the incubator off.
6. Press and hold the inverted Y button on the RH module
and turn the power switch on. Press the inverted Y key
until the beeping stops.
The incubator RH is ready for calibration when the tem-
perature display and CO
2
displays read 00.0. If the word SET
appears in the LCD window, press the inverted y button to
remove it.
When in the RH Calibration mode, the display shows the
temperature (°C) of the water in the humidity reservoir, not the
RH concentration.
7a. Using the arrow touch buttons, adjust the display
upward or downward. If the RH reading was lower than
the reference instrument, increase the display by 0.2°C
for each percent that the (RH) incubator reading was
low.
Example:
If the RH display reads 85% and the reference
instrument reads 87%, increase the display reading by 0.4°C.
7b. If the RH Control reading was higher than the reference
instrument, decrease the display by 0.2°C for each per-
cent that the (RH) incubator reading was high.
Example:
If the RH display reads 87% and the reference
instrument reads 85%, decrease display by 0.4°C.
8. Turn the incubator off, then back on without pressing
any buttons.
9. Return the Program/Run key switch to Run.
c. Temperature Calibration
Any change in temperature calibration will
require adjustments to the temperature
alarm/monitor, RH control and temperature con-
trol. Failure to make these adjustments will result
in incorrect humidity control, poor temperature
uniformity, or both.
Equipment Required: a thermometer of known accuracy.
1. Place the reference thermometer in the center of the
chamber.
2. Close both incubator doors and allow the incubator tem-
perature to stabilize for about one hour.
3. Record the temperature indicated on the thermometer
and the temperature readout on the Temperature
Alarm/Monitor Module.
4. Carefully swing open the control panel and identify the
Temperature Alarm module.
5. Locate the adjustment
potentiometer labeled
Zero. Refer to the illustra-
tion at the right.
6. Using the adjustment
screwdriver, turn the Zero
pot until the displayed
temperature agrees with
the reference thermometer.
7. Close the control panel,
taking care not to pinch
the ribbon wires inside the
compartment.
8. Turn the Program/Run key switch to Program.
9. Turn the incubator off.
10. Press and hold the inverted Y touch button on the RH
module while turning the power switch on. Press the
inverted Y button until the beeping stops. When the unit
is ready for calibration, the temperature display and the
CO
2
display will read 00.0.
Note:
If the word SET appears in the lcd window, momentarily
press the inverted Y button to remove it.
11. Using the up or down arrow buttons, change the display
by an amount equal to the amount the Temperature
Alarm/Monitor display was changed.
Example:
If the reference instrument reads 37.5 and the
Temperature Alarm/Monitor reads 37.0, the display would have
been increased 0.5. Therefore, the RH Control display must
also be increased 0.5.
12. Turn the incubator off.
13. Press and hold the inverted Y touch button on the
Temperature Module while turning the incubator on.
Note:
Continue to press the inverted Y button until the beeping
stops. when the unit is ready for calibration, the RH display and
the CO
2
display will read 00.0. If the word Set appears in the
display window, press the inverted Y button to remove it.
Model 3033 and 3860 Series _________________________________________________________________Service
5 - 2
Figure 5-1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 3033
Page 6: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 8: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 33: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 37: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 41: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 45: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...