2.17 Connecting to the Remote Alarm Contacts
An internal DPDT relay is provided on the Temperature
Alarm/Monitor to monitor alarms and is connected to an RJ-11
(telephone style) jack on the back of the cabinet. Figure 2-10
identifies the pin contacts, Figure 2-1 shows the location of the
Remote Alarm Connector.
Figure 2-11
The remote contacts are “dry” contacts, rated at
30V, 1A maximum.
A modular to modular cable (Forma Stock No.
190388) and an RJ-11 telephone style terminal con-
verter (Stock No. 190392) or equivalent may be used
to convert the remote alarm output to a screw terminal
connection. Refer to Figure 2-12.
Model 3033 and 3860 Series ____________________________________________________Installation and Start-Up
2 - 6
Figure 2-12
2.18 Convenience/Accessory Outlet
An electrical outlet is located on the back of the incubator,
providing 115VAC 0.35A max for lights and accessories. The
maximum allowable leakage current for the product plugged
into the outlet is 3 mA. The outlet is protected by an adjacent
fuse 0.5 amp, 250V, T (time-lag), Forma P/N 230120. Refer to
the electrical schematic in Section 9 of this manual.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 3033
Page 6: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 8: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 33: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 37: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 41: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 45: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...