Working Area Requirements
Sample Fume Extraction
Thermo Scientific
iCE 3000 Series Pre-Installation Requirements Guide (P/N 9499 500 30031, Version 1.6)
5-15
shows the extraction positions of iCE 3000 Series
instruments.
Sample Fume Extraction
Additional separate extraction should be considered, if significant
numbers of volatile or acidic samples are left in the proximity of the
instrumentation, even those in an autosampler.
Figure 5-2.
Extraction positions iCE 3000 Series
250 mm
511 mm
150 mm
566 mm
588 mm
362 mm
150 to
230 mm
594 mm
788 mm
Æ
250
Æ
150
Æ
80
Æ
110
360 mm
657 m
m
529 mm
The extraction system must not be affected by external weather
conditions or other equipment connected to the same exhaust system.