Working Area Requirements
Spectrometer Size
5-6
iCE 3000 Series Pre-Installation Requirements Guide (P/N 9499 500 30031, Version 1.6)
Thermo Scientific
Multiple installations require separate mains supplies.
Additional standard mains sockets are required for the Data Station and
the printer, for an extended autosampler, for a chiller and for each
additional accessory.
Each electrical outlet must have an effective earth/ground connection.
This protection must not be negated by using an extension cable
without a protective earth conductor.
Spectrometer Size
lists the dimensions and weights for the iCE 3000 Series AA
spectrometers.
The values for CETAC ASX-260 and CETAC ASX-520 given in
are based on CETAC specifications for the ASX-520 family.
Table 5-3.
Spectrometer dimensions and weights
Spectrometer
Width
Depth
Height
Weight
iCE 3500
780 mm
(30.1 in)
577 mm
(22.7 in)
520 mm
(20.5 in)
54 kg
(119 lbs)
iCE 3400
780 mm
(30.1 in)
577 mm
(22.7 in)
520 mm
(20.5 in)
54 kg
(119 lbs)
iCE 3300
567 mm
(22.3 in)
589 mm
(23.2 in)
510 mm
(20.1 in)
45 kg
(99 lbs)
iCE 3300GF
567 mm
(22.3 in)
589 mm
(23.2 in)
510 mm
(20.1 in)
38 kg
(84 lbs)
Table 5-4.
Accessory dimensions and weights
Accessory
Width
Depth
Height
Weight
Furnace
265 mm
(10.4 in)
495 mm
(19.5 in)
390 mm
(15.4 in)
50 kg
(110.2 lbs)
Zeeman
Furnace
265 mm
(10.4 in)
495 mm
(19.5 in)
390 mm
(15.4 in)
66 kg
(145.5 lbs)
Furnace
Autosampler
296 mm
(11.7 in)
320 mm
(12.6 in)
380 mm
(15 in)
7.3 kg
(16.1 lbs)
EC100
305 mm
(12 in)
500 mm
(19.7 in)
215 mm
(8.5 in)
16 kg
(35.3 lbs)
VP100
340 mm
(13.4 in)
360 mm
(14.2 in)
280 mm
(11 in)
8.5 kg
(18.7 lbs)
ID100
200 mm
(7.9 in)
200 mm
(7.9 in)
170 mm
(6.7 in)
3.1 kg
(6.8 lbs)