ARL EQUINOX 100
Page
21
SERVICE!
Behind the bigger right side cover, you will find two rings with grip to maintain the
gas cylinder.
1- Access for gas cylinder replacement and
cooling circuit refilling.
2 - Access for X-Ray source alignment.
WARNING!
PRESSURIZED GAS:
If the gas regulator is not provided with the instrument, it is the responsibility of the
customer to install his own gas regulator and, on the input gas circuitry, a safety
valve preloaded at max 6.5 Bar to avoid any over pressure inside of the instrument.
6.4 Connection Specifications
6.4.1 Computer Communication
The ARL EQUINOX is shipped with a computer that already has all software required for
operating the instrument. You just have to connect the RJ45 cable to the back panel of the
ARL EQUINOX X-Ray diffractometer and to the Ethernet port of the computer. The RJ45
cable used is a patch cable (straight).