One Lambda, Inc. | A Thermo Fisher Scientific Brand
Micro SSP Gel System User Manual
TDX-OLI-DMR-PS-0619 Rev. 02
6
Figure 3: Dowel Holes of the Gel Box
5. Lock the Gel Box onto the Base by sliding the locking pin into the single dowel hole located on the
black end of the Gel Box. Once in the locked position and tighten the thumb screw snug. Ensure that
the Gel Box is fastened securely to the Base.
6. Adjust the height of the three legs on the Base to center the leveling bubble; this assures a level
surface (figure 4). The adjustment thumbscrews should not be disturbed after it has been leveled.
Figure 4: Leveling bubble
7. Orient and fully insert all 14 Gel Combs (2 Electrode Combs and 12 Well Combs) into the Gel Comb
Holder (figure 5), this assures the proper alignment of all the Gel Combs and agarose wells during
the gel formation process. If your application does not require 12 rows of wells, you may insert fewer
than 12 well combs, as desired. However, you must always insert the two Electrode Combs in their
respective positions at the ends of the Gel Comb Holder.
Figure 5: Inserting Gel Combs into Gel Comb Holder