3.2 Speed and Time Settings

Speed setting

1. Press RPM/TIME key. The speed 
    indicator light is lit.
2. Press up and down arrow keys to 
    access the speed setting state, speed 
    value starts flashing.
3. Press up and down arrow keys to
    change the speed value. Speed 
    increases or decreases. The minimum 
    unit is 1rpm.
4. During settings, the unit will reset to 
    current values if no value is input after 
    3 seconds.
5. Press “Run/Stop” key to run or stop.

Time setting

1. Press RPM/TIME key. The time 
    indicator light is lit.
2. Press up and down arrow keys to 
    access the time setting state. Speed 
    value starts flashing;
3. Press up and down arrow keys to 
    change the time value. Time increases 
    or decreases. The minimum unit is 1 
    minute or 1 hour.
4. During settings, the unit will reset to 
    current values if no value is input after 
    3 seconds.
5. Press “Run/Stop” key to run or stop.

Notes
1. When the shaker is running, the LED 
    display will switch between the speed 
    and timer function every 30 seconds 

    automatically. Press RPM/TIME key to 
    manually toggle between the speed 
    and timer value.
2. In “timer mode” the LED display will 
    show the remaining running time.
3. In “continuous mode” the LED display 
    will show elapsed run time.
4. After the run is complete, the LED 
    display will read “end”. Press any key 
    to place the shaker in standby mode.

3

.

3

 

Alarm System

Alarm System

1. During operation, the alarm will sound 
    after 10 seconds if the unit has not 
    begun shaking. In addition, the alarm 
    will sound and stop for the speed over 
    650 rpm after 2 seconds.
2. Once the shaker begins to alarm, the 
    RUN indicator light will not be visible 
    and the WARN indicator light will flash, 
    and the unit will stop running.

Power failure recovery

When the power supply is cut off 
suddenly to the shaker, resulting in a 
power failure, the unit will automatically 
run at the previous set parameter upon 
power restoration.

Section 4  Application

Application on Lab Benches

To operate the shaker outside of a CO2 incubator, refer to Specifications in Section 2.

Application in CO2 Incubator

Place the main body of the shaker in the CO2 incubator, and use the 
signal cable to connect the control box through the access port of the CO2
incubator. Refer to Figure 4-1.

Warning

1. Never put the control box in the CO2 
    incubator.
2. The back of control box is magnetized, 
    and can be attached on the surface of 
    the CO2 incubator.
3. When the main body of CO2 Resistant 
    Shaker is in operation in the CO2 
    incubator, keep a safe distance from 
    the inner wall of the CO2 incubator to 
    avoid collisions.

4. The main body of the shaker should 
    be connected to the control box wire 
    through the access port of the CO2 
    incubator.
5. If the cells are sensitive to vibrations, 
    do not put them in the same incubator 
    or on the shaker.

11        Section 3  |  Operation

Figure 4

-

1

1) CO2 Incubator

2) Control Box

3) Power Cord

4) Main body of CO2 Resistant Shaker

MaxQ 2000 CO2 Plus      12

 

 88883101

Summary of Contents for MaxQ 2000 CO2 Plus

Page 1: ...r online thermofisher ca For Research Use Only Not for use in diagnostic procedures 2020 Thermo Fisher Scientific Inc All rights reserved All trademarks are the property of Thermo Fisher Scientific an...

Page 2: ...xtended Warranty for your Thermo Scientific products Whatever Thermo Scientific products you need or use we will be happy to discuss your applications If you are experiencing technical problems workin...

Page 3: ...al Cable Output current of Control Box Maximum Shaking Speed Section 3 Operation Operations on the Control Panel Speed and Time Settings Alarm System Section 4 Application Section 5 Safety Tips Sectio...

Page 4: ...xQ 2000 CO2 Plus Promotion Package for EMEA APAC includes CO2 resistant shaker 88881102B Universal Platform Flask Clamp Starter Kit Main Body Control Box General Power Adaptor US Plug CN Plug EU Plug...

Page 5: ...recessed countersink head screws 2 18inch aluminum platform 3 Main body of CO2 resistant shaker 3 Section 1 Unpacking and Installation 1 2 Universal Aluminum Platform Installation 1 4 Rubber Mat Plat...

Page 6: ...00 m Humidity 20 to 99 non condensing 2 2 Environmental Conditions Rotation Speed Load Timer Dimensions Weight Power Supply Others 5 Section 2 Overview 2 3 Safety Tips 1 Unplug and unload before movin...

Page 7: ...1000ml 1000ml 250 2000ml 2000ml 200 3000ml 3000ml 200 5000ml 5000ml 150 Flask size Size Liquid volume Max Speed rpm 50ml 50ml 600 600 600 600 600 580 550 500 470 430 450 400 450 400 100ml 100ml 150ml...

Page 8: ...has been designed with a soft start slow down providing a smooth acceleration deceleration during operation This feature will decrease splashing of solutions from vessels during operation Warning The...

Page 9: ...operation the alarm will sound after 10 seconds if the unit has not begun shaking In addition the alarm will sound and stop for the speed over 650 rpm after 2 seconds 2 Once the shaker begins to alarm...

Page 10: ...e power cord 2 Remove the platform with 4 screws 3 Remove top plate with 7 screws 4 Use 70 ETOH to clean the inner area 5 All allow the unit to dry for one hour reassemble unit and operate normally Se...

Page 11: ...6mm 1 88881122B 18inch Aluminum platform N A 360x305x10 88881123 Rubber Mat platform 10 130x90x15 88881124 96 well microplate holder Table 6 2 18inch Aluminum Platform Description 50ml clamp 100ml cla...

Page 12: ...Accessories Section 7 Troubleshooting To troubleshoot the unit refer to the table below If your problem persists contact your local sales service representative Indicator off No reaction No rotating...

Page 13: ...the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo...