Herasafe 2030i / Maxisafe 2030i
50155215 |
51
Operation |
Chapter 6
Startup Mode
Figure 35 shows the Main Screen while the unit is starting up. The following startup
screen shows what happens when the “AutoStart”
or “Fan” button is tapped: The
screen shows alerts for the air barrier because the airflow speeds are not up to speed yet.
Figure 35
Main Screen - Unit starting up
The symbol graph shows that the window is opened to the working position, and the
animated yellow arrow and air barrier symbols indicate that the airflow is not ready. The
three status indicator boxes show:
a green checkmark icon in the window status box to indicate that the window is
ready, but
yellow “Timer” icons with rotating hands show in the Downflow and Inflow boxes,
signifying that the airflows are not up to proper speed yet.
Figure 35 shows a typical startup mode example for “AutoStart”, with the “Lights”, “Fans”,
and “Power receptacle” buttons already enabled (see “Programs” on page 86 for more
details). If enabled in AutoStart, the “Gas” button
may come on later when the airflows
are up to speed.
Standby Mode
Figure 36 shows the Main Screen in standby mode.
Figure 36
Main Screen - Unit in Standby
NOTE When the operational downflow and inflow speeds are reached, the
yellow timers are replaced by green checkmark icons.
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...