background image

 
 

Doc No.  

November 2018 

 

4 mm Chromatography                   

2 mm Chromatography 

P/N: 045877                                      

P/N: 045878 

1 Coil   = 1.5 

 3.0 mL/min              

1 Coil = 0.37 

 0.75 mL/min         

2 Coils = 0.5 

 1.5 mL/min              

2 Coils = 0.12 

 0.37 mL/min         

 

P

3

, ________ psi 

 

Calculate the backpressure across the backpressure coil(s) (

BP

2

):  

             

BP

2

, = 

P

3

 

 

P

2

 = ________ psi                                                    

Pass:

 30-40 psi for carbonate eluent or 10-40 psi for other eluents 

 
If the back pressure exceeds the specification, see 

Backpressure Coil Adjustment

 

section 
  

 

 

II.

 

Hydration & Backpressure Check of Suppressor 

Before putting the new suppressor to use, hydrate it properly first. Improper hydration 
will result in irreversible damage to suppressor. 

This will void the warranty!

 

 

Step 4:

 Connect the 

column out 

tubing to the suppressor 

Eluent

 

Inlet

 and direct 

the suppressor 

Eluent Outlet

 to the 

Regen Inlet

 of the suppressor using ~ 

6”

 of 

green  PEEK  tubing 

(ID  0.030”)

  provided  with  the  system  ship  kit.  Direct  the 

suppressor 

Regen Outlet

 to waste (see 

Figure 1

 below).  

 

T

urn on the pump and flow ≤10

 mM eluent or DI water at 

0.5 mL/min

 for 4mm 

suppressor or 

0.125 mL/min

 for 2mm suppressor for 20 minutes. 

The power to 

the suppressor must be turned off during this step.

  

 

 

 

 

Figure 1:

 

Suppressor Hydration Configuration

 

Reviews: