- 30 -
Installation
and
Operation
Reservoir Overflow Device
The overflow device is designed to act as an automatic drain for the unit's
reservoir. The assembly is stainless steel with the exception of the drain fitting
which is Ryton
®
. The unit can still be drained as normal, but if the level reaches
the middle of the tee it will begin to overflow out the bent tube.
Drain the unit's reservoir.
Remove and discard the stop screw from the
bottom of the unit's drain housing.
Remove the unit's drain fitting.
Thread the overflow assembly into the drain
housing. (You can remove the vertical tube
assembly from the Swagelok
®
fitting if
there is a clearance issue.)
After threading the overflow assembly
in as far as it goes, loosen it back out
until the Swagelok
®
fitting is at the top.
Install one of the two supplied 4-40 x
¼ phillips head screws into the
bottom of the drain housing where the
stop screw use to be. (Make sure this
screw is tight so that the assembly
doesnt turn and drain the tank.)
Reinstall the vertical tube assembly into the Swagelok
®
if it was removed.
Install the drain fitting into the end of the new assembly.
Install the other supplied 4-40 x ¼"phillips head screw into the bottom of the
overflow assembly. (This prevents the drain fitting from falling off the unit.)
Description
Part Number
RTE 7 and RTE 10 overflow device
126000000044
RTE 17 overflow device
126000000045
RTE 740 overflow device
126000000049
Drain Housing
Overflow Assembly
Drain Fitting
Swagelok
®
Fitting
Vertical Tube