7
START
Is
Setpoint
< actual
?
Calibration
error
NO
YE S
Replace
valve
Valve is
mechanically OK
Test
STOP
Refer to
ELECTRICAL & CONTROL
section
Rem ove coil
from valve
Drawing 67.cfl
Rick M ills
13 March 1995
YE S
YE S
NO
Valve is
stuck open
Not a
Refrigeration
Problem
Problem is in
Controller or
sensor
See VALVE
REPLACEMENT
D oes
unit
warm up?
YE S
NO
UNSURE
YE S
NO
NO
Can you
get cool
lamp off? *
Is
coo l lamp
on?
* by m anipulating setpoint
Is
coil
on?
No rmal
Response
Troubleshooting Chart: Runaway cold