G. Frequently Asked Questions (FAQ)
the low-pressure neutron detector would be identified as either a GNid
+L +C +BT or a GN +L +C +BT.
Q.
Will Reach-Back work with all wireless telephones?
A.
No... In order to be compatible with Reach-Back, the wireless phone
must have Bluetooth® technology, an internal modem, and dial-up
networking (DUN) in order to access a Reach-Back server.
Q.
Should the device be returned once periodically for calibration?
A.
ICx Radiation, recommends returning the InterceptorTM annually for
optimization and once every 2 years to have the battery replaced.
Q.
Can spectra be downloaded from the Interceptor TM using the
winTMCA32 software?
A.
No... this software package is used for further analysis of spectra.
Q.
What do the functional test results indicate?
A.
See chart below:
Functional Test
Description
High Voltage Test
Verifies the detector’s voltage is
within tolerance
Spectra Test
No test
Device Communication Test
Checks the communications
during the test command
Battery Communication Test
Checks the battery monitor
communications which is
where the Battery Gas Gauge
percentage is derived
Battery Voltage Test
Checks the battery voltage to
make sure it is between the
Max (4.6V) and Min (2.9V)
limits
to be continued. . .
Thermo Fisher Scientific
Interceptor™/en/5.1(4956)/Feb2010
139
Summary of Contents for Interceptor
Page 1: ...en 5 1 4956 02 2010 User Manual Interceptor Spectroscopic Personal Radiation Detector ...
Page 2: ...B Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 3: ...Interceptor Spectroscopic Personal Radiation Detector User Manual ...
Page 36: ...20 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 122: ...106 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 132: ...116 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 142: ...126 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 144: ...128 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 146: ...130 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 148: ...132 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 169: ...I Extended Nuclide Library Thermo Fisher Scientific Interceptor en 5 1 4956 Feb2010 153 ...