B. Specifications
. . . continuation
Explosive Atmospheres
UL913-2002
The I
™ does not comply
with UL913-2002. It is not recom-
mended that the I
™ be
used in environments where an intrin-
sically safe instrument is required.
Thermo Fisher Scientific Corporation
does not assume any liability arising
from the use of the I
™ in
an environment where an intrinsically
safe instrument is required.
Controls
Keys
3 sealed, light touch switches with
excellent clear, tactile feel
Display
Alarm Indicators
Red LED (gamma)
Blue LED (neutron)
Sound
Vibration (10000
±
2000 rpm, 0.8g)
Power Status Indicator
Green LED
Normal
Blinking 11 times a minute,
0.5 sec. ON, 5 sec. OFF
Low Battery
Blinking 50 times a minute,
0.2 sec. ON, 1 sec. OFF
Charging
Blinking 8 times a minute,
2.5 sec. ON, 5 sec. OFF
Charged
Blinking,
0.5 sec. ON, 0.5 sec. OFF
to be continued. . .
Thermo Fisher Scientific
Interceptor™/en/5.1(4956)/Feb2010
123
Summary of Contents for Interceptor
Page 1: ...en 5 1 4956 02 2010 User Manual Interceptor Spectroscopic Personal Radiation Detector ...
Page 2: ...B Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 3: ...Interceptor Spectroscopic Personal Radiation Detector User Manual ...
Page 36: ...20 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 122: ...106 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 132: ...116 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 142: ...126 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 144: ...128 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 146: ...130 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 148: ...132 Interceptor en 5 1 4956 Feb2010 ICx Technologies ...
Page 169: ...I Extended Nuclide Library Thermo Fisher Scientific Interceptor en 5 1 4956 Feb2010 153 ...