Universal Chromatography Interfaces
UCI-50 and UCI-100
Operating Instructions
Page
41
Message
Probable Cause
Remedial action
Error reading from UCI-100 @
USB-00-E0-EF-DE-XX-XX
Data error (cyclic redundancy
check)
Communication error between the
UCI and the Chromeleon server
PC.
Check the USB connection. The
connection to the next hub must
not exceed 5 m. The overall
connection length, including the
hub connections must not exceed
30 m.
Replace any defective USB cable
or hub.
Error reading from UCI-100 @
USB-00-E0-EF-DE-XX-XX
The connection between the UCI
and the Chromeleon server may be
interrupted.
Check the USB connection.
The power supply to the UCI may
be interrupted.
Check the connection to the power
supply of the UCI.