1
TriPlus RSH Site Preparation
Environment Requirements
10
TriPlus RSH Preinstallation Requirements Guide
Thermo Scientific
Environment Requirements
The operating environment in your laboratory is affected by such factors as temperature,
humidity, particulate matter, and electrostatic discharge.
It is your responsibility to provide an acceptable operating environment for your TriPlus RSH.
Attention to the operating environment will ensure continued high performance of your
TriPlus RSH.
Environmental Conditions
• Internal use
• Up to 3000 meters altitude over sea level
• Temperature 5 to 40 °C (41 to 104 °F)
• Maximum relative humidity 80%, non-condensing
• Voltage variations must not exceed the nominal voltage by ± 10%
• Transient overloads in compliance with installation categories II
• Pollution degree according to IEC 664 (3.7.3) 2
• Protection degree IP00
Use the following guidelines to ensure your site has the proper environmental conditions for
the system:
1. Ensure that your room temperature is 5-40 °C (41-104 °F). The analytical performance is
only confirmed for temperatures between 15-35 °C (59-95 °F). For best performance, the
operating temperature should be constant. Use the table below to calculate the amount of
heat your system will generate and ensure your air-conditioning system can handle that
amount of heat.
DSQ II MS
700
ITQ MS
700
TSQ-Quantum XLS MS
2310
Computer *
400
Monitor *
25
* Power requirements vary by manufacturer.
Table 7.
System Power Requirements (Sheet 2 of 2)
Equipment
Maximum Power (W)