Chapter 15
PID Troubleshooting
Maintaining a Photoionization Detector
Maintenance and Troubleshooting Manual
335
Troubleshooting the PID
Table 15-3.
PID troubleshooting table
Symptom
Diagnosis
Remedy
UV Lamp does not light
immediately
The lamp has not been used
for a long time.
Set
High current mode?
to
Y
until
the lamp lights, the reset to
N
UV Lamp does not light
at all.
Message:
Lamp failure
Lamp not installed
Install a suitable UV lamp.
Refer to the
Operating Sequence on page 319.
Lamp not correctly installed
in the lamp holder
Reinstall the UV lamp.
Refer to the
Operating Sequence on page 319.
Lamp holder not correctly
fixed to the lamp housing
Tighten the two knurled screws that
fix the lamp holder to the housing.
Lamp power cable
disconnected
Check the connection
Polarization cable
disconnected
Check the connection
Safety interlock circuit is
faulty
Contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
UV lamp is exhausted
Replace UV lamp.
Refer to the
Operating Sequence on page 319.
No standing current
Power and signal cables not
properly connected
Check the connections of the cables
Tailing of solvent peak
Insufficient make-up gas flow
rate
Increase the flow rate.