Chapter 15
Replace the UV Lamp
Maintaining a Photoionization Detector
Maintenance and Troubleshooting Manual
319
OPERATING SEQUENCE
Replace the UV Lamp
Materials needed
•
UV Lamp (see Table 15-2 on page page 318)
•
Viton
TM
O-ring for UV lamp (P/N 290 203 38)
•
Cotton gloves
WARNING! UV radiation can cause injuries to your eyes. Therefore, before beginning the
sequence, make sure the UV lamp has been switched off.
PRECAUTIONS
WARNING! The lamp replacement can be carried out at operating temperature, without
removing the detector from its base body. Nevertheless, keep in mind that parts of
the detector are hot and could cause injuries to your hands.
PRECAUTIONS