Chapter 9
Maintaining a PTV Injector
PTV Troubleshooting
172
Maintenance and Troubleshooting Manual
Troubleshooting the PTV Injector
Table 9-3.
PTV Troubleshooting Table
Symptom
Diagnosis
Remedy
Discrimination of
heavy compounds in
splitless mode
Transfer temperature too low
Set
Transfer temp
to a higher
value.
Transfer time too low
Set
transfer time
or
cleaning
time
to a value close to the whole GC
run time.
Splitless time too short
Set
Splitless time
to a higher
value.
Initial temperature too high
with regard to the solvent
boiling point
Set
Inject temp
to a value closer to
the solvent boiling point.
Flooding of the liner due to
low injection temperature
Set
Inject temp
to a higher value. If
the symptom does not disappear, insert
quartz wool in the liner. Refer to
Operating
Sequence on page 165.
Flooding of the liner due to
injected volume too large
Reduce the amount of sample to be
injected.
Replace the liner with one of adequate
diameter. Refer to
Operating Sequence on page 151.
Flooding of the liner due to
insufficient size
Replace the liner with one of larger
diameter. Refer to
Operating Sequence on page 151.
Dirty liner
Clean or replace the liner. Refer to
Sequence on page 151.