Chapter 9
Clean or Replace the Liner
Maintaining a PTV Injector
Maintenance and Troubleshooting Manual
151
OPERATING SEQUENCE
Clean or Replace the Liner
Materials needed
•
Screwdriver (P/N 205 026 03)
•
Liner (see Table 9-2, on page 145)
•
Graphite seal (P/N 290 134 17)
•
Ultrasonic bath
•
Methanol/acetone mixture (1:1)
•
Non metallic sharp tool
•
Septum (P/N 313 132 25)
•
Tweezers
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the injector must be cooled to room temperature.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.
1. While in stand-by condition, press
OVEN
to
access the oven control table.
.
OVEN
Temp
50
50 <
Initial time
5
Ramp 1
10.0
2. Scroll to
Temp
and press
OFF
.